List of parameters for hart communicator – Yokogawa ADMAG AXF User Manual
Page 100
![background image](/manuals/360071/100/background.png)
IM 1E10C1-E
11-8
11. PARAMETER LIST
List of Parameters for HART Communicator
PV
–32400 to 32400
R
Displays instantaneous flow rate in engineering unit.
A20
Totl
0 to 999999
0
W
Displays and presets forward direction totalized flow rate.
A30
PV AO
2.40 to 21.6 mA
R
Displays current output.
—
Span
0.0001 to 30000
1.0000
W
Sets flow span in selected unit.
b03
Status group 1
Error
R
Displays self-diagnostics result.
—
µ
P fault
“ON” shows error status.
EEPROM fault
“OFF” shows normal status.
A/D fault
See “8.2 Self-diagnostics Functions.”
Signal overflow
Coil open
Vel. span> 10 m/s
Vel. span< 0.3 m/s
Status group 2
P. span> 1000p/s
R
Ditto
—
P. span> 500p/s
P. span> 25p/s
P. span> 15p/s
P. span> 10p/s
P. span> 5p/s
P. span< .0001p/s
T. span> 1000p/s
Status group 3
T. span< .0001p/s
R
Ditto
—
Analog lmt error
Empty pipe
Multi range err
Coil short
EEPROM no set
EEPROM default
Dev id not entered
Self test
—
—
W
Excecutes self-diagnostics.
—
Auto zero
—
—
W
Executes automatic zero adjustment.
C02
Zero tuning
(00) Inhibit
Enable
W
Inhibits or enables automatic zero adjustment.
C01
(01) Enable
D/A trim
ZERO: 3.2 to 5.6 mA
W
Performs fine adjustments of zero and span of current output.
—
SPAN: 18.4 to 21.6 mA
Scaled D/A trim
ZERO: 3.2 to 5.6 mA
W
Performs fine adjustments of zero and span of scaled analog
—
SPAN: 18.4 to 21.6 mA
output.
Loop test
4 mA
4 mA
W
Sets test output value.
H02
20 mA
Other (2.72 to 21.28 mA)
End
Status out
(00) Normal
Normal
W
Selects test status output.
H03
(01) Closed (on)
(02) Open (off)
PV Unit
gal/min
m/s
W
Selects engineering unit of “PV”. Displays “Spcl” in case of
—
L/min
setting unit except 5 units on the left.
Cum/h
ft/s
m/s
(Spcl)
Name
Data Range, Units
Default
Value
R/W
Description
Indicator
of SE