beautypg.com

Adp300/adp305 test record, Adp30x probe, Equipment used – Teledyne LeCroy ADP300_305 User Manual

Page 30

background image

ADP30x Probe

ADP300/ADP305 Test Record

Serial Number:

Asset or Tracking Number:

Date:

Technician:


Equipment
Used:

MODEL

SERIAL NUMBER

CALIBRATION

DUE DATE

OSCILLOSCOPE

DIGITAL MULTIMETER

FUNCTION
GENERATOR

1

1The function generator used in this Performance Verification Procedure is used for making relative
measurements. The output of the generator is measured with a DMM or oscilloscope in this
procedure. Thus, the generator is not required to be calibrated.

Step

Description

Intermediate
Data

Test Result

Gain Accuracy

5

Function Generator Output
Voltage

____________V

11

Probe Output Voltage

____________V

13

÷1000 Gain Error (Test limit ±1%)

___________ %

16

Probe Output Voltage

____________V

18

÷100 Gain Error (Test limit ±2%)

___________ %

26

923079-00 Rev A