Voltage Derating for ADP30x Standard Accessories
Accessory (Teledyne LeCroy P/N)
Combined Probe and Accessory
Max Input Voltage (from either input to
ground)
Plunger Hook Clips (PK30X-1)
1000V CAT II (600V CAT III)
Safety Alligator Clips (PK30X-2)
1000V CAT III
Plunger Jaw Clips (PK30X-3)
1000V CAT II (600V CAT III)
Plunger Clamp Clips (PK30X-4)
1000V CAT III
Safety Spade Terminals (PK30X-5)
1000V CAT II (600V CAT III)
NOTE: The operating altitude of ADP30x is derated to 2000 m (6560 ft) when
used with these accessories. The voltage rating and CAT rating are de-rated to
the voltage in the above table when used with these accessories.
WARNING. To avoid risk of electric shock or fire, do not exceed either
the voltage rating or category rating of the probe or the probe
accessory, whichever is the lesser of the two.
WARNING. To avoid risk of electric shock when using the probe or
accessories, keep your fingers behind the finger guard of the probe.
Keep the output cable away from the circuits being measured. Use only the
specified accessories.