beautypg.com

Operation, Connecting the probe to the test instrument, Connecting the probe to the test circuit – Teledyne LeCroy ADP300_305 User Manual

Page 15

background image

Operator’s Manual

Operation

Connecting the Probe to the Test Instrument

The ADP30x series of probes has been designed for use with Teledyne LeCroy

oscilloscopes equipped with the ProBus interface. When you attach the probe

output connector to the scope’s input connector, the scope will recognize the
probe, set the scope input termination to 1 MΩ, and activate the probe control

functions in the user interface.

Connecting the Probe to the Test Circuit

The ADP30x series of probes is ideally suited for applications where no ground is

available or where the location of a ground connection is unknown.
Two inputs are available at the probe tip to connect the probe to a circuit under

test. For accurate measurements, the + and – inputs both must always be

connected to the test circuit. Positive voltages applied to the + input (red)

relative to the – input (blue) will deflect the oscilloscope trace toward the top of

the screen.
To maintain the high performance capability of the probe in measurement

applications, care must be exercised in connecting the probe to the test circuit.

Increasing the parasitic capacitance or inductance in the input paths may

introduce a “ring” or may slow the rise time of fast signals. Input leads that form

a large loop area will pick up any radiated electromagnetic field that passes

through the loop and may induce noise into the probe inputs. Because this signal

will appear as a differential mode signal, the probe’s common mode rejection

will not remove it. This effect can be greatly reduced by twisting the input leads

together to minimize the loop area.
High common mode rejection requires precise matching of the relative gain or

attenuation in the + and – input signal paths. Mismatches in additional parasitic

capacitance, inductance, delay, and a source impedance difference between the

+ and – signals will lower the common mode rejection ratio. Therefore, it is

desirable to use the same length and type of wire and connectors for both input

connections. When possible, try to connect the inputs to points in the circuit

with approximately the same source impedance.

923079-00 Rev A

11