Ransburg, Needle / electrode – Ransburg Vector R Series Cascade 79523 R90 Waterborne User Manual
Page 42
Vector R Series Cascade Applicators - Maintenance
Ransburg
38
Reinstall
1. Check the needle/electrode tightness on the
needle shaft. If it is loose, tighten it (refer to "Nee-
dle/Electrode" in the "Maintenance" section).
2. With the applicator trigger actuated, place the
fluid nozzle over the needle/electrode and screw
it into the barrel by hand.
3. Tighten it using the special multi-purpose
wrench with 3/8" square. Torque to 40-45 lbs - in
(4.5-5.1 Nm). An alternate method (if no torque
wrench is available) is to hand tighten, then tight-
en an additional 1/16 to 1/8 turn in the clockwise
direction.
4. Install the air cap and retaining ring onto the
applicator (refer to "Air Cap - Reinstall" in the
"Maintenance" section).
Figure 20: Reinstalling Fluid Nozzle
DO NOT over-tighten the fluid nozzle
into the barrel. Doing so could damage
or break the fluid nozzle or damage the
thread of the barrel.
NOTE
NEEDLE / ELECTRODE
Removal
1. Remove the air cap and fluid nozzle from the
applicator assembly.
2. Secure the needle shaft at the rear of the barrel
and unscrew the needle/electrode from the needle
shaft.
Cleaning and Inspection
1. Use a suitable solvent to clean the needle/
electrode.
2. Examine the needle/electrode for damage or
wear. Pay special attention to the area where
the wire electrode extends from the main body.
This is a sealing surface that seats inside the fluid
nozzle. If there are signs of wear in this area,
both the needle/electrode and fluid nozzle must
be replaced.
3. An electrical check of the needle/electrode must
be done prior to reinstalling it into the applicator
assembly (refer to "Needle/Electrode Resistance
Testing" in the "Maintenance" section).
Reinstall
1. Secure the needle shaft at the rear of the barrel
and screw the needle/electrode into place by hand.
2. Reinstall the fluid nozzle and air cap onto the
barrel.
C W
3 5 -4 0
L B S -I N
3 . 9 -4 . 5 N M
AH-06-01.13