Appendix

A-1

Specifications

Packaging: General Purpose

• Flush panel mount (Standard).

• Relay rack mount. Contains one

instrument in one 19" relay rack
mountable plate (Optional).

Sensor: Teledyne BDS Sensor, patent pending.

Sample System: All wetted parts of 316 stainless steel with

built-in restrictor.

90 % Response Time: Less than 90 seconds at 25 °C (77 °F) on

10, and 100 ppm range. 90 seconds on
1000ppb range.

Software programmable response in 100
ppb range from 1 minute to 60 minutes.
Default is 5 minutes response time.

Ranges: Three user definable ranges from

0–100 ppb to 0–100 ppm, plus over range
of 0-100 ppm.

Autoranging with range ID output.

Alarms: One system-failure alarm contact to detect

power failure or sensor-zero and span
failure.

Two adjustable concentration threshold
alarm contacts with fully programmable
setpoints.

Displays: 2-line by 20-character, VFD screen, and

one 5 digit LED display.