Teledyne BDS-3000 - Oxygen analyzer User Manual
Page 39
Oxygen Analyzer
Installation
Teledyne Analytical Instruments
25
System Alarm:
Actuates when DC power supplied to circuits is
unacceptable in one or more parameters. Permanently
configured as failsafe and latching. Cannot be defeated.
Actuates if self-test fails.
(Reset by pressing button to remove power. Then
press again and any other button EXCEPT System to
resume.
Further detail can be found in Chapter 4, Section 4-5.
Table 3-2: Alarm Relay Contact Pins
Pin
Contact
45
Threshold Alarm 1, normally closed contact
28
Threshold Alarm 1, moving contact
46
Threshold Alarm 1, normally open contact
42
Threshold Alarm 2, normally closed contact
44
Threshold Alarm 2, moving contact
43
Threshold Alarm 2, normally open contact
36
System Alarm, normally closed contact
20
System Alarm, moving contact
37
System Alarm, normally open contact
Digital Remote Cal Inputs: Accept 0 V (off) or 24
V dc (on) inputs for
remote control of calibration. (See Remote Calibration Protocol below.)
See Table 3-3 for pin connections.
Zero:
Floating input. 5 to 24 V i nput across the + and – pins puts
the analyzer into the Zero mode. Either side may be
grounded at the source of t he signal. 0 to 1 volt across the
ter minals allows Zero mode to ter minate when done. A
synchronous signal must open and close the external zero
val ve appropriately. See Remote P robe Connect or. (T he –C
opt ion internal valves oper ate automati cally) .