Oxygen Analyzer

Installation

Teledyne Analytical Instruments

25

System Alarm:

Actuates when DC power supplied to circuits is
unacceptable in one or more parameters. Permanently
configured as failsafe and latching. Cannot be defeated.
Actuates if self-test fails.

(Reset by pressing button to remove power. Then
press again and any other button EXCEPT System to
resume.

Further detail can be found in Chapter 4, Section 4-5.

Table 3-2: Alarm Relay Contact Pins

Pin

Contact

45

Threshold Alarm 1, normally closed contact

28

Threshold Alarm 1, moving contact

46

Threshold Alarm 1, normally open contact

42

Threshold Alarm 2, normally closed contact

44

Threshold Alarm 2, moving contact

43

Threshold Alarm 2, normally open contact

36

System Alarm, normally closed contact

20

System Alarm, moving contact

37

System Alarm, normally open contact

Digital Remote Cal Inputs: Accept 0 V (off) or 24

V dc (on) inputs for

remote control of calibration. (See Remote Calibration Protocol below.)
See Table 3-3 for pin connections.

Zero:

Floating input. 5 to 24 V i nput across the + and – pins puts
the analyzer into the Zero mode. Either side may be
grounded at the source of t he signal. 0 to 1 volt across the
ter minals allows Zero mode to ter minate when done. A
synchronous signal must open and close the external zero
val ve appropriately. See Remote P robe Connect or. (T he –C
opt ion internal valves oper ate automati cally) .