beautypg.com

Teledyne BDS-3000 - Oxygen analyzer User Manual

Page 65

background image

Oxygen Analyzer

Operation

Teledyne Analytical Instruments

51

#### PPM

Zero

Slope = ####

ppm/s

The beginning zero level is shown in the upper left corner of the

display. As the zero reading settles, the screen displays and updates
information on slope (unless the slope starts within the acceptable zero
range and does not need to settle further).

Then, and whenever slope is less than 0.08 for at least 3 minutes,

instead of slope you will see a countdown: 5 Left, 4 Left, and so fourth.
These are five steps in the zeroing process that the system must
complete, AFTER settling, before it can go back to Analyze.

#### PPM

Zero

4 Left = ###

ppm/s

The zeroing process will automatically conclude when the output is

within the acceptable range for a good zero. Then the analyzer
automatically returns to the Analyze mode.

Because the reading of the slope is not very sensitive, it is

recommended that zero gas be purging a few minutes before starting the
Auto mode zeroing. This will ensure cell stability on the new Zero
settings.

4.4.1.2 M

ANUAL

M

ODE

Z

EROING

Press ZERO to enter the Zero function. The screen that appears

allows you to select between automatic or manual zero calibration. Use
the __

keys to toggle between

AUTO

and

MAN

zero settling. Stop when

MAN

appears, blinking, on the display.

Zero: Settling: Man
To Begin

Press ENTER to begin the zero calibration. After a few seconds the

first of five zeroing screens appears. The number in the upper left-hand
corner is the first-stage zero offset. The microprocessor samples the
output at a predetermined rate. It calculates the differences between
successive samplings and displays the rate of change as slope= a value
in parts per million per second (ppm/s).

#### ppm

Zero

Slope = ####

ppm/s

Note: It takes several seconds for the true slope value to display.

Wait about 10 seconds. Then, wait until Slope is