Oxygen Analyzer

Operation

Teledyne Analytical Instruments

53

Use the __

arrow keys to enter the oxygen-concentration mode

(999.99 ppm is maximum value of span gas allowed). Use the__arrow
keys to blink the digit you are going to modify. Use the __

arrow keys

again to change the value of the selected digit. When you have finished
typing in the concentration of the span gas you are using, press ENTER
to begin the Span calibration.

#### ppm Span
Slope = #### ppm/s

The beginning span value is shown in the upper left corner of the

display. As the span reading settles, the screen displays and updates
information on slope. Spanning automatically ends when the span output
corresponds, within tolerance, to the value of the span gas concentration.
Then the instrument automatically returns to the analyze mode.

4.4.2.2 M

ANUAL

M

ODE

S

PANNING

Press SPAN to start the Span function. The screen that appears

allows you to select whether the span calibration is to be performed
automatically or manually.

Span: Settling: MAN
For Next

Use the __

keys to toggle between

AUTO

and

MAN

span settling.

Stop when

MAN

appears, blinking, on the display. Press ENTER to

move to the next screen.

Press ENTER to move to the next screen.

Calib. Holding time
Cal hold: 5 min

This menu allows the operator to set the time the analyzer should

be held in the

AUTO

span mode. It does not have any effect in the

MAN

ual mode. Just press ENTER key to continue.

Span Val:

008.00ppm

Span Mod

#

Press _() to permit modification (

Mod #

) of span value.

Use the arrow keys to enter the oxygen concentration of the span

gas you are using (999.99 is maximum value of span gas). The__ arrows
choose the digit, and the __arrows choose the value of the digit.