Teledyne BDS-3000 - Oxygen analyzer User Manual

Page 74

background image

Operation

BDS 3000

Teledyne Analytical Instruments

60

ppm O

2

Output (VDC)

Output (mA DC)

0

0.0

4.0

10

0.1

5.6

20

0.2

7.2

30

0.3

8.8

40

0.4

10.4

50

0.5

12.0

60

0.6

13.6

70

0.7

15.2

80

0.8

16.8

90

0.9

18.4

100

1.0

20.0

The analog output signal has a voltage which depends on the

oxygen concentration AND the currently activated analysis range. To
relate the signal output to the actual concentration, it is necessary to
know what range the instrument is currently on, especially when the
analyzer is in the autoranging mode.

To provide an indication of the range, a second pair of analog

output terminals are used. They generate a steady preset voltage (or
current when using the current outputs) to represent a particular range.
Table 4-3 gives the range ID output for each analysis range.