beautypg.com

4 calibration of the analyzer – Teledyne BDS-3000 - Oxygen analyzer User Manual

Page 63

background image

Oxygen Analyzer

Operation

Teledyne Analytical Instruments

49

0.00 ppb/degree C

The number on the second line is adjustable using the Up and

Down arrow keys between 0.00 and 10.0. The coefficient is different
from sensor to sensor and it’s entered into the analyzer at the factory
before shipping. If the sensor is replaced, a new coefficient must be
entered. TAI can give the coefficient or it may be estimated.

To estimate it in the field:

1. Set the coefficient to zero.

2. Run the analyzer on “Zero” calibration gas for two weeks or

once stability of its baseline is reached.

3. After the sensor has been purged for at least two weeks and it’s

baseline is stable, monitor the reading and ambient temperature
over a minimum period of 24 hours.
Take the maximum and minimum readings, and the maximum
and minimum temperature readings.

4. Calculate the coefficient using the relation:

Coefficient = (O

2

max - O

2

min) ÷ (Temp max - Temp min)

For example:

In a 24 hour run:

O

2

max = 3.55 ppb

O

2

min = 1.75 ppb

Temp max = 24.5 degrees C.

Temp min = 22.1 degrees C.

Coefficient = (3.55-1.75)ppb ÷ (24.5-22.1) °C = 0.75 °C.

4.4

Calibration of the Analyzer

The analyzer must be calibrated prior to its use. For most

applications where the desired range of measurement is 0 to 10 ppm, or
less we recommend the analyzer be calibrated using a span gas with a
concentration between 7.0 to 9.0 ppm oxygen in nitrogen. This will
require that calibration be performed in the 0-10 ppm analyzer range.

Before the cell is ready for calibration, it must be purged with

sample gas to a low oxygen level—preferably below 0.1 ppm. If the