Oxygen Analyzer

Operation

Teledyne Analytical Instruments

39

CAUTION:

FREQUENT ZERO ADJUSTMENTS OF THE CELL IS
NOT RECOMMENDED. A NEWLY INSTALLED CELL
MAY TAKE 7-10 DAYS OF OPERATION TO REACH A
STEADY ZERO (TYPICALLY LESS THAN 5 PPB). IF
REQUIRED, THE INSTRUMENT MAY BE ZEROED
AFTER THIS INITIAL STABILIZING PERIOD AND MAY
BE CHECKED AGAIN AFTER AN ADDITIONAL 7-10
DAY PERIOD. THE FREQUENCY OF ZERO
ADJUSTMENT IS AT THE DISCRETION OF THE
USER (ONCE A MONTH IS SUGGESTED).

Choose System from the Function buttons. The LCD will display

five subfunctions.

TRAK/HLD Auto—Cal
PSWD Logout More

Use < > arrows to blink

Auto—Cal

, and press Enter. A new screen

for Span/Zero set appears.

Span OFF Nxt: 0d 0h
Zero OFF Nxt: 0d 0h

Press < > arrows to blink

Span

(or

Zero

), then press ENTER again.

(You won’t be able to set OFF to ON if a zero interval is entered.) A

Span Every

... (or

Zero Every

...) screen appears.

Span Every 0 d
Start 0 h from now

Use __ arrows to set an interval value, then use __ arrows to move

to the start-time value. Use DÑ arrows to set a start-time value.

To turn ON the Span and/or Zero cycles (to activate Auto-Cal):

Press System again, choose

Auto—Cal,

and press ENTER again. When

the Span/Zero values screen appears, use the < > arrows to blink the
Span (or Zero) OFF/ON field. Use _ arrows to set the OFF/ON field to
ON. You can now turn these fields ON because there is a nonzero span
interval defined.

4.3.3

Password Protection

If a password is assigned, then setting the following system

parameters can be done only after the password is entered:

span

and

zero

settings,

alarm

setpoints, analysis

range

definitions, switching between

autoranging

and manual override, setting up an

auto-cal

, and assigning a