Oxygen Analyzer
Installation
Teledyne Analytical Instruments
27
For example:
1) Test the CRC. When the CRC is open, Send a zero command
until the CRC closes (The CRC will quickly close.)
2) When the CRC closes, remove the zero command.
3) When CRC opens again, send a span command until the
CRC closes. (The CRC will quickly close.)
4) When the CRC closes, remove the span command.
When CRC opens again, zero and span are done, and the sample is
being analyzed.
Note: The Remote Valve connections (described below) provides
signals to ensure that the zero and span gas valves will be
controlled synchronously.
Range ID Relays: Four dedicated Range ID relay contacts. The first
three ranges are assigned to relays in ascending order—Low range is
assigned to Range 1 ID, Medium range is assigned to Range 2 ID, and
High range is assigned to Range 3 ID. The fourth range is reserved for
the Air Cal Range (25%). Table 3-4 lists the pin connections.
Table 3-4: Range ID Relay Connections
Pin
Function
21
Range 1 ID Contact
38
Range 1 ID Contact
22
Range 2 ID Contact
39
Range 2 ID Contact
19
Range 3 ID Contact
18
Range 3 ID Contact
34
Range 4 ID Contact (Air Cal)
35
Range 4 ID Contact (Air Cal)
Network I/O: A serial digital input/output for local network protocol.
At this printing, this port is not yet functional. It is to be used for future
options to the instrument. Pins 13 (+) and 29 (–).