beautypg.com

Teledyne BDS-3000 - Oxygen analyzer User Manual

Page 40

background image

Installation

BDS 3000

Teledyne Analytical Instruments

26

Span:

Floating input. 5 to 24 V input across the + and – pins
puts the analyzer into the Span mode. Either side may be
grounded at the source of the signal. 0 to 1 volt across the
terminals allows Span mode to terminate when done. A
synchronous signal must open and close external span
valve appropriately. See Figure 3-5 Remote Probe
Connector. (The –C option internal valves operate
automatically.)

Cal Contact:

This relay contact is closed while analyzer is spanning
and/or zeroing. (See Remote Calibration Protocol below.)

Table 3-3: Remote Calibration Connections

Pin

Function

9

+ Remote Zero

11

– Remote Zero

10

+ Remote Span

12

– Remote Span

40

Cal Contact

41

Cal Contact

Remote Calibration Protocol: To properly time the Digital Remote Cal
Inputs to the Model BDS 3000 Analyzer, the customer's controller must
monitor the Cal Relay Contact.

When the contact is OPEN, the analyzer is analyzing, the Remote

Cal Inputs are being polled, and a zero or span command can be sent.

When the contact is CLOSED, the analyzer is already calibrating. It

will ignore your request to calibrate, and it will not remember that
request.

Once a zero or span command is sent, and acknowledged (contact

closes), release it. If the command is continued until after the zero or
span is complete, the calibration will repeat and the Cal Relay Contact
(CRC) will close again.