Introduction

BDS 3000

Teledyne Analytical Instruments

6

Alarm Connections

2 concentration alarms and 1
system alarm.

RS-232 Port

Serial digital concentration signal
output and control input.

Remote Probe

Used in the BDS 3000 for
controlling external solenoid
valves only.

Remote Span/Zero

Digital inputs allow external
control of analyzer calibration.

Calibration Contact

To notify external equipment that
instrument is being calibrated and
readings are not monitoring
sample.

Range ID Contacts

Four separate, dedicated, range
relay contacts. Low, Medium,
High, Cal.

Network I/O

Serial digital communications for
local network access. For future
expansion. Not implemented at this
printing.

Note: If you require highly accurate Auto-Cal timing, use external

Auto-Cal control where possible. The internal clock in the
Model BDS 3000 is accurate to 2-3 %. Accordingly,
internally scheduled calibrations can vary 2-3 % per day.