Maintenance

BDS 3000

Teledyne Analytical Instruments

66

Figure 5.1 Adding water into the BDS sensor

WARNING:

THE SENSOR USED IN THE MODEL BDS 3000
OXYGEN ANALYZER USES ELECTROLYTE WHICH
CONTAINS POTASSIUM HYDROXIDE, THAT CAN BE
HARMFUL IF TOUCHED, SWALLOWED, OR
INHALED. AVOID CONTACT WITH ANY FLUID OR
POWDER IN OR AROUND THE UNIT. WHAT MAY
APPEAR TO BE PLAIN WATER COULD BE THE
ELECTROLYTE. IN CASE OF EYE CONTACT,
IMMEDIATELY FLUSH EYES WITH WATER FOR AT
LEAST 15 MINUTES. CALL PHYSICIAN. (SEE
APPENDIX, MATERIAL SAFETY DATA SHEET.)

5.3

Fuse Replacement

1. Place small screwdriver in notch, and pry cover off, as shown in

Figure 5-2.

2. To change between American and European fuses, remove the

single retaining screw, flip Fuse Block over 180 degrees, and
replace screw.

3. Replace fuse as shown in Figure 5-3.

4. Reassemble Housing as shown in Figure 5-2.