beautypg.com

Teledyne BDS-3000 - Oxygen analyzer User Manual

Page 57

background image

Oxygen Analyzer

Operation

Teledyne Analytical Instruments

43

4.3.4

Logout

The Logout function provides a convenient means of leaving the

analyzer in a password protected mode without having to shut the
instrument off. By entering

Logout

, you effectively log off the

instrument leaving the system protected against use until the password is
reentered. To log out, press the System button to enter the System
function.

TRAK/HLD Auto—Cal
PSWD Logout More

Use the __ arrow keys to position the blinking over the

Logout

function, and press ENTER to Log out. The screen will display the
message:

Protected Until
Password Reentered

4.3.5

System Self-Diagnostic Test

The Model BDS 3000 has a built-in self-diagnostic testing routine.

Pre-programmed signals are sent through the power supply, output board
and sensor circuit. The return signal is analyzed, and at the end of the
test the status of each function is displayed on the screen, either as OK
or as a number between 1 and 3. (See System Self Diagnostic Test in
Chapter 5 for number code.)

The self diagnostics are run automatically by the analyzer

whenever the instrument is turned on, but the test can also be run by the
operator at will. To initiate a self diagnostic test during operation:

Press the System button to start the System function.

TRAK/HLD Auto—Cal
PSWD Logout More

Use the < > arrow keys to blink More, then press Enter.

Version Diag

Xout

Neg-Y Filter-10

More

Use the __arrow keys again to move the blinking to the Self–Test

function. The screen will follow the running of the diagnostic.

RUNNING DIAGNOSTIC
Testing Preamp — 83