Liquid Controls HML110 IOM User Manual

Page 8

background image


HML 110

8

110_EN_LC_5_3_6X.doc


INTERNAL VIEW OF CONVERTER












CONVERTER TO SENSOR ELECTRICAL CONNECTIONS











Power supply

1

2

3

4

5

6

7

8

9

10

11

121

314

15

1617

1819

20

90..265VAC

>HV<

15...45VAC
10...63VDC

>LV<

Dangerous voltage on block 12-13:
- 60 Vdc Max

- 250 V Max on commutation coils (switching)

Signalling LED:
See interpretation notes page 12

IF2 socket

Power supply

Keyboard

C018 CABLE

1
2

3
4

11
12

13

E1
E2

C
SH

B1

B2

1

2

3

4

5

6

7

8

9

10

11

12

13

14

15

16

17

18

19

20

-

E C

+

SC

B2

B1

SC

OUT2

OUT1

24V

24V

COILS

E1

E2

C

SC

4-20mA

ELECTRODES

E C

+

-

B

A

RS 485*

INPUT

-

+

RS 485* : OPTIONAL

SEPARATE VERSION

TERMINAL BLOCK

Sudden movements to the electrodes
cable can cause noise during

measurement Max length of

electrodes cable: 20 m