Cleaning and decontamination, Troubleshooting, Warning list – Hach-Lange POLYMETRON 9586 User Manual
Page 22: Replacement parts and accessories

Cleaning and decontamination
D A N G E R
Always remove power from the controller before performing maintenance activities.
The analyzer does not normally require any cleaning or decontamination. If needed, clean the
exterior of the instrument with a moist cloth and a mild soap solution. Never use cleaning agents
such as turpentine, acetone or similar products to clean the instrument, including the display and any
accessories.
Troubleshooting
Warning list
A warning icon consists of an exclamation point within a triangle. Warning icons appear on the right
of the main display below the measurement value. A warning does not affect the operation of menus,
relays and outputs. To view warnings, push the menu key and select DIAGNOSTICS. Then select
the device to view any problems associated with that device. The warning icon will no longer be
displayed once the problem has been corrected or acknowledged.
A list of possible warnings is shown in
Table 5 Warning list for oxygen scavenger sensors
Warning
Description
Resolution
O2 SCAVENGER TOO
HIGH
The measured value is >
10000 ppm
Make sure that the oxygen scavenger
(hydrazine or carbohydrazide) level in the
process water is within the operating limits of
the sensor. Calibrate or replace the sensor.
O2 SCAVENGER TOO LOW The measured value is < 0 ppm
Calibrate or replace the sensor.
TEMP TOO HIGH
The measured temperature is >
50 °C
Reduce sample temperature.
TEMP TOO LOW
The measured temperature is <
0 °C
Increase sample temperature.
CURRENT TOO HIGH
The measured current > 200 mA Make sure that the oxygen scavenger
(hydrazine or carbohydrazide) level in the
process water is within the operating limits of
the sensor. Calibrate or replace the sensor.
CURRENT TOO LOW
The measured current < -0.5 mA Calibrate or replace the sensor.
CAL OVERDUE
The Cal Reminder time has
expired
Calibrate the sensor.
REPLACE SENSOR
The sensor has been in
operation > 365 days
Replace the sensor cartridge and calibrate the
sensor. If the calibration result is pass, reset
the membrane days in the DIAG/TEST menu.
NOT CALIBRATED
The sensor has not been
calibrated
Calibrate the sensor.
CAL IN PROGRESS
A calibration was started but not
completed
Return to calibration.
Replacement parts and accessories
Refer to the replacement parts and accessories section of the controller documentation for controller
parts and accessories.