Page 16

background image

3.4.3 Density

measurement


The attainable accuracy depends on the selected calibration type.

Without calibration no density measurement is possible and the empty pipe recognition
is not available!

Model

without

3-Point

5-Point

CMU-C

5 g/l

2 g/l

CMU-D

5 g/l

2 g/l

CMU-E

5 g/l

1 g/l

CMU-G

5 g/l

1 g/l

CMU-H

5 g/l

1 g/l

CMU-J

5 g/l

2 g/l

CMU-K

5 g/l

2 g/l

CMU-L

5 g/l

2 g/l

CMU-N

5 g/l

2 g/l

CMU-P

5 g/l

2 g/l

CMU-Q

5 g/l

2 g/l

CMU-R

5 g/l

2 g/l

no mea

surement

of density

Density accuracy

3.4.4 Accuracy

Mass flow

Accuracy CMU-C to CMU-H

± 0.1% of actual flow + zero point stability (see Section 3.4.2 CMU
flow ranges)

Accuracy CMU-J to CMU-R

± 0.15% of actual flow + zero point stability (see Section 3.4.2 CMU
flow ranges)

Repeatability error

± 0.05% of actual flow (sensor with transmitter)

Additional measured values

Volume flow

± 0.2 % of actual value + zero point stability

Temperature

± 0.5 °C

Hysteresis n/a
Settling time

1 to 15 seconds

Startup drift

15 minutes

Long-term drift

± 0.02 % of upper-range value per year

Influence of ambient temperature

± 0.005 % per K

Influence of fluid temperature

Compensated

Influence of fluid pressure

For fluids: too small to be relevant


Page 16 of 112