Section 4 Maintenance

Installation and Operation Manual

X-TMF-5850i-MFC-eng

Part Number: 541B108AAG

December, 2008

Example: What is the air equivalent flow rate of 2000 sccm Hydrogen?

In order to calculate the orifice conversion factor when using a gas mixture,
the following formula must be used:

Where P1

=

percentage by volume of gas 1

P2

=

percentage by volume of gas 2

Pn

=

percentage by volume of gas n

Example: Find the Air equivalent flow for 20 slpm of a 20% Helium
and 80% Chlorine gas mixture.

Q

AIR

=

Q

gas

(orifice conversion factor)

=

20 x 1.417

=

28.34 slpm air

2. If inlet and outlet pressures are given in gauge pressure (psig)

add14.7 to convert to absolute pressure (psia).

Outlet Pressure — 30 psig + 14.7 = 44.7 psia
Inlet Pressure — 50 psig + 14.7 = 64.7 psia

3. Determine Critical Pressure Drop

Critical pressure drop occurs when the outlet pressure (psia) is less
than half the inlet pressure (psia) or

P outlet

<

P inlet

2

If these conditions exist, the pressure drop (

Δp) should be calculated as

follows: