Brooks, Model 5850i – Brooks Instrument 5850i User Manual

Page 14

background image

2-4

Brooks

®

Model 5850i

Section 2 Installation

Installation and Operation Manual

X-TMF-5850i-MFC-eng

Part Number: 541B108AAG

December, 2008

Note: The control valve in the Model 5850i provides precision control and
is not designed for positive shut-off. If positive shut-off is required, it is
recommended that a separate shut-off valve be installed in-line.

2-6 In-Line Filter

It is recommended that an in-line filter be installed upstream from the
controller to prevent the possibility of any foreign material entering the flow
sensor or control valve. The filtering element should be replaced
periodically or ultrasonically cleaned.

Maximum Flow Rate

Recommended Filter Size

100 sccm

1 micron

500 sccm

2 microns

1 to 5 slpm

7 microns

10 to 30 slpm

15 microns

Figure 2-2 "D" Type Connector Pin Arrangement

1

SETPOINT RETURN

9

SUPPLY COMMON

2

VOLTAGE SIGNAL OUTPUT

10

SIGNAL OUTPUT RETURN

3

NOT USED

11

5V REFERENCE OUTPUT

4

CURRENT SIGNAL OUTPUT

12

VALVE OVERRIDE INPUT

5

+15 TO +28 Vdc SUPPLY

13

NOT USED

6

NOT USED

14

CHASSIS GROUND

7

CURRENT SETPOINT INPUT

15

NOT USED

8

VOLTAGE SETPOINT INPUT

*JUMPER SELECTABLE

NOTE: Cable shield tied to chassis ground in connector. Make no
connection on customer end.

Table 2-1 Recommended Filter Size