Dynasonics TFXL Clamp-On Ultrasonic Flow User Manual
Page 22

22
06-TTM-UM-00158 8/2012
are examined by the Bad Data Rejection fi lter. The value is 
entered as a percentage of actual fl ow rate.
For example, if the average fl ow rate is 100 GPM and the 
Flow Filter Hysteresis is set to 5%, a fi lter window of 95-105 
GPM is established. Successive fl ow measurements that are 
measured within that window are recorded and averaged in 
accordance with the Flow Filter Damping setting. Flow read-
ings outside of the window are held up in accordance with 
the Bad Data Rejection fi lter.
Flow Filter MinHysteresis sets a minimum hysteresis 
window that is invoked at sub 0.25 FPS (0.08 MPS) fl ow rates, 
where the “of rate” Flow Filter Hysteresis is very small and 
ineff ective. This value is entered in pico-seconds (ρ sec) and 
is diff erential time. If very small fl uid velocities are to be 
measured, increasing the Flow Filter MinHysteresis value 
can increase reading stability.
Flow Filter Sensitivity allows confi guration of how fast the 
Flow Filter Damping will adapt in the positive direction. 
Increasing this value allows greater damping to occur faster 
than lower values. Adaptation in the negative direction is not 
user adjustable.
Bad Data Rejection is a value related to the number of 
successive readings that must be measured outside of 
the Flow Filter Hysteresis or Flow Filter MinHysteresis 
windows before the fl ow meter will use that fl ow value. 
Larger values are entered into Bad Data Rejection when 
measuring liquids that contain gas bubbles, as the gas 
bubbles tend to disturb the ultrasonic signals and cause 
more extraneous fl ow readings to occur. Larger Bad Data 
Rejection values tend to make the fl ow meter more sluggish 
to rapid changes in actual fl ow rate.
OUTPUT TAB
The entries made in the Output tab establish input and 
output parameters for the fl ow meter. Select the appropriate 
function from the pull-down menu and press the Down-
load button. When a function is changed from the factory 
setting, a Confi guration error (1002) will result. This error 
will be cleared by resetting the TFXL microprocessor from 
the Communications/Commands/Reset Target button or 
by cycling power on the TFXL fl ow meter. Once the proper 
output is selected and the microprocessor is reset, calibration 
and confi guration of the modules can be completed.
FIGURE 4.6 - OUTPUT TAB
CHANNEL 1 - 4-20 mA FREQUENCY 
CONFIGURATION
NOTE: The 4-20 mA Output Frequency Menu applies to all 
TFXL versions and is the only output choice for Channel 1.
The Channel 1 menu controls how the 4-20 mA output is 
spanned for all TFXL models.
The Flow at 4 mA / 0 Hz and Flow at 20 mA / 1,000 Hz settings 
are used to set the span for both the 4-20 mA output and the 
0-1,000 Hz frequency output on the TFXL meter versions.
The 4-20 mA output is internally powered (current sourcing) 
and can span negative to positive fl ow rates. This output 
interfaces with virtually all recording and logging systems 
by transmitting an analog current that is proportional to 
system fl ow rate. Independent 4 mA and 20 mA span settings 
are established in fi rmware using the fl ow measuring range 
entries. These entries can be set anywhere in the - 40 to + 40 
FPS (-12 to +12 MPS) range of the instrument. Resolution of 
the output is 12-bits (4096 discrete points) and can drive up 
to a 900 Ohm load. When powered by a DC supply, the load is 
limited by the input voltage supplied to the instrument. See 
Figure 3.1 for allowable loop loads.
Flow at 4 mA / 0 Hz
Flow at 20 mA / 1,000 Hz
The Flow at 4 mA / 0 Hz and Flow at 20 mA / 1,000 Hz entries 
are used to set the span of the 4-20 mA analog output and 
the frequency output on TFXL versions. These entries are 
volumetric rate units that are equal to the volumetric units 
confi gured as rate units and rate interval discussed on 
Page 23.
For example, to span the 4-20 mA output from -100 GPM to 
+100 GPM with 12 mA being 0 GPM, set the Flow at 4 mA / 0 
Hz and Flow at 20 mA / 1,000 Hz inputs as follows:
