beautypg.com

Dynasonics TFXL Clamp-On Ultrasonic Flow User Manual

Page 12

background image

12

06-TTM-UM-00158 8/2012

iterative process. Table 2.2 contains recommended mounting
confi gurations for common applications. These recom-
mended confi gurations may need to be modifi ed for specifi c
applications if such things as aeration, suspended solids, out
of round piping or poor piping conditions are present. Use of
the TFXL diagnostics in determining the optimum transducer
mounting is covered later in this section.

Size

Frequency

Setting

Transducer

Mounting

Mode

½

2 MHz

DTTSnP

V

DTTSnC

DTTSnT

¾

2 MHz

DTTSnP

DTTSnC

DTTSnT

1

2 MHz

DTTSnP

DTTSnC

DTTSnT

2 MHz

DTTSnP

DTTSnC

DTTSnT

2 MHz

DTTSnP

DTTSnC

DTTSnT

2

1 MHz

DTTSnP

DTTSnC

2 MHz

DTTSnT

NOTE: DTTS transducer designation refers to both DTTS
and DTTC transducer types.

TABLE 2.3 - TRANSDUCER MOUNTING

MODES — DTTS / DTTC

STEP 3 - ENTERING PIPE AND LIQUID DATA

The TFXL system calculates proper transducer spacing by
utilizing piping and liquid information entered by the user.
This information can be entered on a TFXL via the software
utility.

The best accuracy is achieved when transducer spacing is
exactly what the TFXL calculates, so the calculated spacing
should be used if signal strength is satisfactory. If the pipe
is not round, the wall thickness not correct or the actual
liquid being measured has a diff erent sound speed than the
liquid programmed into the transmitter, the spacing can vary
from the calculated value. If that is the case, the transducers
should be placed at the highest signal level observed by
moving the transducers slowly around the mount area.

NOTE: Transducer spacing is calculated on “ideal” pipe.
Ideal pipe is almost never found so the transducer spacing
distances may need to be altered. An eff ective way to maxi-

mize signal strength is to confi gure the display to show signal
strength, fi x one transducer on the pipe and then starting
at the calculated spacing, move the remaining transducer
small distances forward and back to fi nd the maximum signal
strength point.

Important! Enter all of the data on this list, save the data and
reset the TFXL before mounting transducers.

The following information is required before programming
the instrument:

Transducer mounting
confi guration

Pipe O.D. (outside diameter)

Pipe wall thickness

Pipe material

Pipe sound speed

1

Pipe relative roughness

1

Pipe liner thickness (if present)

Pipe liner material (if present)

Fluid type

Fluid sound speed

1

Fluid viscosity

1

Fluid specifi c gravity

1

NOTE: Much of the data relating to material sound speed,
viscosity and specifi c gravity is pre-programmed into the TFXL
fl ow meter. This data only needs to be modifi ed if it is known
that a particular application’s data varies from the reference
values. Refer to Part 4 of this manual for instructions on
entering confi guration data into the TFXL fl ow meter via the
software.

1

NOMINAL VALUES FOR THESE PARAMETERS ARE INCLUDED

WITHIN THE TFXL OPERATING SYSTEM. THE NOMINAL
VALUES MAY BE USED AS THEY APPEAR OR MAY BE MODI-
FIED IF EXACT SYSTEM VALUES ARE KNOWN.

After entering the data listed above, the TFXL will calculate
proper transducer spacing for the particular data set. This
distance will be in inches if the TFXL is confi gured in English
units, or millimeters if confi gured in metric units.

STEP 4 - TRANSDUCER MOUNTING

PIPE PREPARATION

After selecting an optimal mounting location (Step 1) and
successfully determining the proper transducer spacing
(Step 2 & 3), the transducers may now be mounted onto the
pipe (Step 4).

Before the transducers are mounted onto the pipe surface,
an area slightly larger than the fl at surface of each transducer
must be cleaned of all rust, scale and moisture. For pipes with
rough surfaces, such as ductile iron pipe, it is recommended
that the pipe surface be wire brushed to a shiny fi nish. Paint
and other coatings, if not fl aked or bubbled, need not be
removed. Plastic pipes typically do not require surface prepa-
ration other than soap and water cleaning.
The DTTN and DTTH transducers must be properly oriented
and spaced on the pipe to provide optimum reliability and
performance. On horizontal pipes, when Z-Mount is required,