2 supplemental characteristics – B&K Precision 9150-9153 - Manual User Manual
Page 47

47
Load Regulation
±(%of output+offset)
Voltage
<0.01% + 0.5 mV
<0.01% + 1 mV
<0.01% + 1 mV
Current
<0.1% + 10 mA
<0.1% + 5 mA
<0.1% + 2 mA
Line Regulation
±(%of output+offset)
Voltage
<0.02% + 0.1 mV
<0.02% + 1 mV
<0.02% + 1 mV
Current
<0.1% + 1 mA
<0.01% + 1 mA
<0.01% + 0.1 mA
Programming resolution
Voltage
0.1 mV
1 mV
1 mV
Current
1 mA
1 mA
0.1 mA
Readback/Meter
resolution
Voltage
0.1 mV
0.1 mV
0.1 mV
Current
1 mA
0.1 mA
0.1 mA
Front Panel setting
resolution
Voltage
0.1 mV
0.5 mV
0.5 mV
Current
1 mA
1 mA
1 mA
Programming accuracy,
12 months (25 °C ± 5 °C)
±(%of output+offset)
Voltage
<0.02%+2 mV
<0.02%+6 mV
<0.02%+12 mV
Current
<0.1%+30 mA
<0.1%+15 mA
<0.05%+10 mA
Readback / Meter
accuracy
12months (25 °C ± 5 °C)
±(%of output+offset)
Voltage
<0.02%+1.5 mV
<0.02%+3 mV
<0.02%+6 mV
Current
<0.05%+15 mA
<0.05%+10 mA
<0.05%+5 mA
Ripple & Noise
(20 Hz ~20 MHz)
Voltage
≤0.005%+3
mVp-p
≤0.005%+3 mVp-p
≤0.005%+4
mVp-p
Current
15 mArms
5 mArms
3 mArms
Temperature coefficient,
(0 °C~40 °C)
±(%of output+offset)
Voltage
<0.02%+2 mV
<0.02%+5 mV
<0.02%+10 mV
Current
<0.1%+30 mA
<0.1%+15 mA
<0.05%+5 mA
Readback temperature,
coefficient,
±(%of output+offset)
Voltage
<0.02%+2 mV
<0.02%+5 mV
<0.02%+10 mV
Current
<0.1%+20 mA
≤0.05%+10 mA
≤0.05%+5 mA
DVM Accuracy
Voltage
0~12 V range: 0.02%+2 mV
0~40 V range: 0.02%+3 mV
DVM Resolution
Voltage
0~12 V range: 0.1 mV
0~40 V range: 1 mV
Ohm meter
m
Ω
Accuracy: 0.1%, for Voltage and Current >= 10% of FS (full
scale)
Accuracy: 0.3% for Voltage and Current = 3 to 10% of FS
Weight
net
63.9 lbs (29 kg)
Dimensions
WxHxD
429 mm × 88.2 mm × 458.9 mm
16.88(W) x 3.47(H) x 18.06(D) inch
*)
LVP: Limit Voltage Protection. Limits the voltage than can be set either via the front panel or remote control command.
NOTE: Specifications and information are subject to change without notice. Please visit
current product information.
5.2 Supplemental Characteristics
State Storage Memory
50 user-configurable stored states
Recommended Calibration Interval