Output commands – B&K Precision 9150-9153 - Manual User Manual
Page 37

37
Parameter: 0~255
Reset value: Consult *PSC command
Example: STATus: OPERation:ENABle 128
Query syntax: STATus: OPERation:ENABle?
Return parameter:
Reference command: *PSC
Output Commands
OUTPut[:STATe]
This command sets the power supply output on or off.
Command syntax: OUTPut[:STATe]
Parameter: 0|1|ON|OFF
*RST value: OFF
Query syntax: OUTPut:STATe?
Return parameter: 0|1
Usage:
If output timer is required, send command OUTPUT:TIMER ON first, then set OUTPUT ON. These two
commands must be set in this order to function properly.
OUTPut:TIMer[:STATe]
This command sets the output timer state of the power supply.
Command syntax: OUTPut:TIMer[:STATe]
Parameter: 0|1|ON|OFF
*RST value: OFF
Query syntax: OUTPut:TIMer[:STATe]?
Return parameter: 0|1
Usage:
To start Timer
-
First, set OUTPUT:TIMER ON, then set OUTPUT ON.
To end Timer
-
Set OUTPUT:TIMER OFF to turn timer off.
To check timer status
-
Send command OUTPUT:TIMER?
OUTPut:TIMer:DATA
This command sets the time of the ou
t
put timer. The unit is in SECOND and decimal fractions cannot be
used for this command.
Command syntax: ONPut:TIMer:DATA
Parameter:
*RST value: 1
Query syntax: OUPut:TIMer:DATA?
Return parameter:
[SOURce:]MODE
This command configures the power supply for command fixed mode, list mode or DVM mode.
FIXed Command fixed mode
LIST List mode
DRM
Digital milliohm meter
Command syntax: [SOURce:]MODE
Parameter: FIXed|LIST|DRM
*RST value: FIXed