B&K Precision 9150-9153 - Manual User Manual
Page 40
40
Parameter: None
Example: LIST:CURR? 1;
Return parameter:
[SOURce:]LIST :VOLTage[:LEVel]
This command sets the voltage step.
Command syntax: [SOURce:]LIST:VOLTage [:LEVel]
Parameter: 0~360V
Unit: V mV
Example: LIST:VOLT 1,3V;
Query syntax: [SOURce:]TRANsition: VOLTage:TLEVel?
Parameter: None
Example: LIST:VOLT? ;
Return parameter:
[SOURce:]LIST:WIDth
This command sets the minimum step time. Decimal fractions are not allowed for this command. Units are in
seconds or milliseconds, which are set using LIST:UNIT command (see LIST:UNIT). Set units first before
using this command.
Command syntax: [SOURce:]LIST:WIDth
Parameter: MIN TO MAX|MIN|MAX
Unit: S mS
Example: LIST:WID 1, 100;
Query syntax: [SOURce:]LIST:WIDth?
Parameter: None
Example: LIST:WID? 1;
Return parameter:
[SOURce:]LIST:NAME
This command sets the name for the list file. Make sure the file name does not exceed 8 characters.
Command syntax: [SOURce:]LIST:NAME
Parameter:
Example: LIST:NAME ‘TEST’;
Query syntax: [SOURce:]LIST:NAME?
Return parameter:
[SOURce:]LIST:AREA
This command divides up the storage area for the list file in one of the 4 ways listed below.
1.1 group per store area, 400 steps
2.2 groups per storage area, each group contains 200 steps.
4.4 groups per storage area, each group has 100 steps.
8.8 groups of storage area, each group has 50 steps.
Command syntax: [SOURce:]LIST:AREA
Parameter: 1|2|4|8
Example: LIST:AREA 1
Query syntax: [SOURce:]LIST:AREA?
Return parameter:
[SOURce:]LIST:SAVe
This command saves the list file to a register (non volatile m
e
mory). The memory can be written
approximately 0.1 million times.
Command syntx:[SOURce:]LIST:SAVe
Parameter: 1~8