Interface configuration commands – B&K Precision 9150-9153 - Manual User Manual
Page 42
42
HIGH: 1W resistance range
Command syntax: SENSe:RESistance:RANGe
Parameter: LOW | MIDdle | HIGH
Example: RES:RANG LOW
Query syntax: SENSe:RESistance:RANGe?
Return parameter:
MEASure[:SCALar]:RESistance[:DC]?
This command queries the resistance reading of the milliohm meter.
Command syntax: MEASure[:SCALar]: RESistance?
Parameter: None
Return parameter:
Return parameter unit: Ohm
Example: MEAS:RES?
Interface Configuration Commands
[SOURce:]SYSTem:SENSe [:STATe]{
This command enables/disables the power supply’s remote sense function.
Command syntax: SYSTem: SENSe [:STATe]
Parameter: 0|1|ON|OFF
Query syntax: SYSTem:SENSe [:STATe]?
*RST value: 0
[SOURce:]PORT:FUNCtion
This command sets the mode of the port in the rear panel.
TRIGGER function:
Pin1, pin2 are configured as external trigger source for the power supply and to
control the list operation.
RI/DFI function:
Inhibit Input controls the output state of the power supply. The Fault Output can
indicate the reason for internal failure.
DIGITAL I/O function: Read and control the state of the digital port
Command syntax: SOURce:PORT:FUNCtion
Parameter: TRIGger|RIDFi|DIGital
Query syntax: SOURce:PORT:FUNCtion?
*RST value: TRIGger
[SOURce:]RI:MODE
This command sets the input mode of the RI input pin
Command syntax: SOURce:RI:MODE
Parameter: OFF|LATChing|LIVE
Query syntax: SOURce:RI:MODE?
*RST value: OFF
[SOURce:]DFI:SOURce
This command sets output source of the DFI output pin. (Not available for models 9150, 9151, 9152, and 9153)
Command syntax: SOURce:DFI:SOURce
Parameter: OFF|QUES|OPER|ESB|RQS
Query syntax: SOURce:DFI:SOURce?
*RST value: OFF