STEP List operation is step mode.
Command syntax: [SOURce:]LIST:MODE
Parameter: CONTinious|STEP
Query syntax: [SOURce:]LIST:MODE?
Return parameter:
[SOURce:]LIST:STEP
This command sets the operation mode of the list file.
ONCE List operate once
REPeat Repeat list operation
Command syntax: [SOURce:]LIST:STEP
Parameter: ONCE|REPeat
Query syntax: [SOURce:]LIST:STEP?
Return parameter:
Note to usage:
When setting to LIST:STEP REPEAT, the list would run repeatedly for an indefinite amount of time.
Below is an example set of commands to setup a repeated list properly:
*IDN?
PORT:FUNCTION TRIGGER
OUTPUT:TIMER OFF
OUPUT OFF
LIST AREA 4
LIST:COUNT 2
LIST:UNIT SECOND
TRIGGER:SOURCE BUS
LIST:VOLTAGE 1,5
LIST:VOLTAGE 2,1
LIST:WIDTH 1,4
LIST:WIDTH 2,2
LIST:MODE CONT
LIST:STEP REPEAT
LIST:NAME ‘MX’
LIST:SAVE 1
MODE LIST
OUTPUT ON
TRIGGER
[SOURce:]LIST:COUNt
This command sets the number of steps for the list operation.
Command syntax: [SOURce:]LIST:COUNt
Parameter: 2~400
Query syntax: [SOURce:]LIST:COUNt?
Parameter: None
Return parameter:
[SOURce:]LIST :CURRent[:LEVel]
This command sets the current step.
Command syntax: [SOURce:]LIST:CURRent[:LEVel]
Parameter: 0~30A
Unit: A mA
Example: LIST:CURR 1,3A;
Query syntax: [SOURce:]LIST:CURRent:[LEVel]?