Dynamic range of the R&S
®
ZVA67
(without optional step attenuators
and without optional direct
generator/receiver access)
from PORT 1 to PORT 2 and
from PORT 3 to PORT 4
10 MHz to 50 MHz
> 70 dB, typ. 90 dB
50 MHz to 500 MHz
> 110 dB, typ. 125 dB
500 MHz to 2 GHz
> 120 dB, typ. 130 dB
2 GHz to 8 GHz
> 130 dB, typ. 140 dB
8 GHz to 12 GHz
> 125 dB, typ. 140 dB
12 GHz to 24 GHz
> 130 dB, typ. 140 dB
24 GHz to 32 GHz
> 120 dB, typ. 130 dB
32 GHz to 40 GHz
> 115 dB, typ. 125 dB
40 GHz to 50 GHz
> 115 dB, typ. 125 dB
50 GHz to 65 GHz
> 107 dB, typ. 115 dB
65 GHz to 67 GHz
> 105 dB, typ. 115 dB
67 GHz to 70 GHz
typ. 103 dB
The dynamic range is defined as the difference between the actually available maximum source power and the RMS value of the
data trace of the transmission magnitude, which is produced by noise and crosstalk with the test ports short-circuited. The
specification is valid without system error correction and at 10 Hz measurement bandwidth. The dynamic range can be increased by
using a measurement bandwidth of 1 Hz. For the R&S
®
ZVA67 at single frequencies below 2.5 GHz, the dynamic range may be
affected by spurious signals.
Dynamic range at optional measurement
input (direct generator/receiver access
option) of the R&S
®
ZVA8
from PORT 1 to MEAS 2 IN
300 kHz to 10 MHz
typ. > 125 dB
10 MHz to 100 MHz
typ. > 135 dB
100 MHz to 8 GHz
typ. > 145 dB
Dynamic range at optional measurement
input (direct generator/receiver access
option) of the R&S
®
ZVA24
from PORT 1 to MEAS 2 IN
10 MHz to 100 MHz
typ. > 135 dB
100 MHz to 13 GHz
typ. > 145 dB
13 GHz to 20 GHz
typ. > 140 dB
20 GHz to 24 GHz
typ. > 130 dB
Dynamic range at optional measurement
input (direct generator/receiver access
option) of the R&S
®
ZVA40
from PORT 1 to MEAS 2 IN
10 MHz to 100 MHz
typ. > 140 dB
100 MHz to 20 GHz
typ. > 150 dB
20 GHz to 24 GHz
typ. > 140 dB
24 GHz to 32 GHz
typ. > 130 dB
32 GHz to 40 GHz
typ. > 120 dB
Dynamic range at optional measurement
input (direct generator/receiver access
option) of the R&S
®
ZVA50
from PORT 1 to MEAS 2 IN
10 MHz to 100 MHz
typ. > 140 dB
100 MHz to 20 GHz
typ. > 150 dB
20 GHz to 24 GHz
typ. > 145 dB
24 GHz to 32 GHz
typ. > 140 dB
32 GHz to 40 GHz
typ. > 135 dB
40 GHz to 50 GHz
typ. > 130 dB
Dynamic range at optional measurement
input (direct generator/receiver access
option) of the R&S
®
ZVA67
from PORT 1 to MEAS 2 IN
10 MHz to 100 MHz
typ. > 140 dB
100 MHz to 20 GHz
typ. > 145 dB
20 GHz to 24 GHz
typ. > 145 dB
24 GHz to 32 GHz
typ. > 140 dB
32 GHz to 40 GHz
typ. > 135 dB
40 GHz to 50 GHz
typ. > 130 dB
50 GHz to 67 GHz
typ. > 125 dB