10
Rohde & Schwarz
R&S
®
ZVA Vector Network Analyzer
Measurement accuracy
This data is valid between +18 °C and +28 °C, provided the temperature has not varied by more than 1 K after calibration. Validity of
the data is conditional on the use of a suitable calibration kit. This calibration kit is used to achieve the effective system data specified
below. Frequency points, measurement bandwidth, and sweep time have to be identical for measurement and calibration (no
interpolation allowed).
Accuracy of transmission measurements
R&S
®
ZVA8
300 kHz to 1 MHz
for +15 dB to –45 dB
< 1 dB or < 6°
1 MHz to 50 MHz
for +15 dB to –30 dB
< 0.2 dB or < 2°
for –30 dB to –45 dB
< 1 dB or < 6°
50 MHz to 8 GHz
for +15 dB to +5 dB
< 0.2 dB or < 2°
for +5 dB to –55 dB
< 0.1 dB or < 1°
for –55 dB to –70 dB
< 0.2 dB or < 2°
for –70 dB to –85 dB
< 1 dB or < 6°
R&S
®
ZVA24
10 MHz to 50 MHz
for +15 dB to –30 dB
< 1 dB or < 6°
50 MHz to 400 MHz
for +15 dB to –30 dB
< 0.2 dB or < 2°
for –30 dB to –45 dB
< 1 dB or < 6°
400 MHz to 700 MHz
for +15 dB to –50 dB
< 0.2 dB or < 2°
for –50 dB to –65 dB
< 1 dB or < 6°
700 MHz to 24 GHz
for +15 dB to +5 dB
< 0.2 dB or < 2°
for +5 dB to –55 dB
< 0.1 dB or < 1°
for –55 dB to –70 dB
< 0.2 dB or < 2°
for –70 dB to –85 dB
< 1 dB or < 6°
R&S
®
ZVA40
10 MHz to 50 MHz
for +15 dB to –30 dB
< 1 dB or < 6°
50 MHz to 250 MHz
for +15 dB to –30 dB
< 0.2 dB or < 2°
for –30 dB to –45 dB
< 1 dB or < 6°
250 MHz to 700 MHz
for +15 dB to +5 dB
< 0.3 dB or < 3°
for +5 dB to –65 dB
< 0.2 dB or < 2°
for –65 dB to –80 dB
< 1 dB or < 6°
700 MHz to 2 GHz
for +15 dB to +5 dB
< 0.3 dB or < 3°
for +5 dB to –50 dB
< 0.1 dB or < 1°
for –50 dB to –65 dB
< 0.2 dB or < 2°
for –65 dB to –80 dB
< 1 dB or < 6°
2 GHz to 24 GHz
for +15 dB to +5 dB
< 0.3 dB or < 3°
for +5 dB to –55 dB
< 0.1 dB or < 1°
for –55 dB to –70 dB
< 0.2 dB or < 2°
for –70 dB to –85 dB
< 1 dB or < 6°
24 GHz to 32 GHz
for +15 dB to +5 dB
< 0.3 dB or < 3°
for +5 dB to –45 dB
< 0.2 dB or < 2°
for –45 dB to –60 dB
< 0.3 dB or < 3°
for –60 dB to –75 dB
< 1 dB or < 6°
32 GHz to 40 GHz
for +15 dB to +5 dB
< 0.4 dB or < 4°
for +5 dB to –40 dB
< 0.2 dB or < 2°
for –40 dB to –55 dB
< 0.4 dB or < 4°
for –55 dB to –70 dB
< 1 dB or < 6°