Specifications, Measurement range – Atec Rohde-Schwarz-ZVA Series User Manual
Page 4
Version 10.00, May 2012
4
Rohde & Schwarz
R&S
®
ZVA Vector Network Analyzer
Specifications
Specifications apply under the following conditions: 90 minutes warm-up time at ambient temperature, specified environmental
conditions met, calibration cycle adhered to, and all internal automatic adjustments performed. "Typical values" are designated with
the abbreviation "typ.". These values are verified during the final test but are not assured by Rohde & Schwarz. "Nominal values" are
design parameters that are not assured by Rohde & Schwarz. These values are verified during product development but are not
specifically tested during production.
Unless otherwise stated, specifications apply to test ports and a nominal source power of –10 dBm.
Measurement range
Impedance
50
Ω
Test port connector
R&S
®
ZVA8
type N, female
R&S
®
ZVA24
3.5 mm, male, ruggedized
R&S
®
ZVA40
2.92 mm, male, ruggedized
R&S
®
ZVA40
2.4 mm, male, ruggedized
R&S
®
ZVA50
2.4 mm, male, ruggedized
R&S
®
ZVA67
1.85 mm, male, ruggedized
Number of test ports
2 or 4
Frequency range
R&S
®
ZVA8
300 kHz to 8 GHz
R&S
®
ZVA24
10 MHz to 24 GHz
R&S
®
ZVA40
10 MHz to 40 GHz
R&S
®
ZVA50
10 MHz to 50 GHz
R&S
®
ZVA67
10 MHz to 67 GHz
Static frequency accuracy
without optional oven quartz
8×10
–6
with optional oven quartz
1×10
–7
Frequency resolution
1 Hz
Number of measurement points
user-selectable
1 to 60001
Measurement bandwidths
1/2/5 steps
1 Hz to 1 MHz
Dynamic range of the R&S
®
ZVA8
(without optional step attenuators
and without optional direct
generator/receiver access)
from PORT 1 to PORT 2 and
from PORT 3 to PORT 4
300 kHz to 50 MHz
> 100 dB, typ. 110 dB
50 MHz to 100 MHz
> 120 dB, typ. 130 dB
100 MHz to 4 GHz
> 130 dB, typ. 140 dB
4 GHz to 7 GHz
> 125 dB, typ. 135 dB
7 GHz to 8 GHz
> 120 dB, typ. 130 dB
Dynamic range of the R&S
®
ZVA24
(without optional step attenuators
and without optional direct
generator/receiver access)
from PORT 1 to PORT 2 and
from PORT 3 to PORT 4
10 MHz to 100 MHz
> 90 dB, typ. 105 dB
100 MHz to 700 MHz
> 105 dB, typ. 120 dB
700 MHz to 2 GHz
> 125 dB, typ. 130 dB
2 GHz to 13 GHz
> 130 dB, typ. 135 dB
13 GHz to 24 GHz
> 125 dB, typ. 130 dB
Dynamic range of the R&S
®
ZVA40
(without optional step attenuators
and without optional direct
generator/receiver access)
from PORT 1 to PORT 2 and
from PORT 3 to PORT 4
10 MHz to 50 MHz
> 90 dB, typ. 100 dB
50 MHz to 500 MHz
> 105 dB, typ. 115 dB
500 MHz to 2 GHz
> 125 dB, typ. 135 dB
2 GHz to 20 GHz
> 130 dB, typ. 140 dB
20 GHz to 24 GHz
> 125 dB, typ. 135 dB
24 GHz to 32 GHz
> 120 dB, typ. 130 dB
32 GHz to 40 GHz
> 118 dB, typ. 125 dB
Dynamic range of the R&S
®
ZVA50
(without optional step attenuators
and without optional direct
generator/receiver access)
from PORT 1 to PORT 2 and
from PORT 3 to PORT 4
10 MHz to 50 MHz
> 90 dB, typ. 100 dB
50 MHz to 500 MHz
> 105 dB, typ. 115 dB
500 MHz to 2 GHz
> 125 dB, typ. 135 dB
2 GHz to 20 GHz
> 130 dB, typ. 140 dB
20 GHz to 24 GHz
> 125 dB, typ. 135 dB
24 GHz to 40 GHz
> 120 dB, typ. 130 dB
40 GHz to 45 GHz
> 115 dB, typ. 125 dB
45 GHz to 50 GHz
> 110 dB, typ. 120 dB