Atec Rohde-Schwarz-ZVA Series User Manual
Page 11
Version 10.00, May 2012
Rohde & Schwarz
R&S
®
ZVA Vector Network Analyzer 11
R&S
®
ZVA50
10 MHz to 50 MHz
for +15 dB to –30 dB
< 1 dB or < 6°
50 MHz to 250 MHz
for +15 dB to –30 dB
< 0.2 dB or < 2°
for –30 dB to –45 dB
< 1 dB or < 6°
250 MHz to 700 MHz
for +15 dB to +5 dB
< 0.3 dB or < 3°
for +5 dB to –65 dB
< 0.2 dB or < 2°
for –65 dB to –80 dB
< 1 dB or < 6°
700 MHz to 2 GHz
for +15 dB to +5 dB
< 0.3 dB or < 3°
for +5 dB to –50 dB
< 0.1 dB or < 1°
for –50 dB to –65 dB
< 0.2 dB or < 2°
for –65 dB to –80 dB
< 1 dB or < 6°
2 GHz to 24 GHz
for +15 dB to +5 dB
< 0.3 dB or < 3°
for +5 dB to –55 dB
< 0.1 dB or < 1°
for –55 dB to –70 dB
< 0.2 dB or < 2°
for –70 dB to –85 dB
< 1 dB or < 6°
24 GHz to 32 GHz
for +15 dB to +5 dB
< 0.3 dB or < 3°
for +5 dB to –45 dB
< 0.2 dB or < 2°
for –45 dB to –60 dB
< 0.3 dB or < 3°
for –60 dB to –75 dB
< 1 dB or < 6°
32 GHz to 40 GHz
for +15 dB to +5 dB
< 0.4 dB or < 4°
for +5 dB to –40 dB
< 0.2 dB or < 2°
for –40 dB to –55 dB
< 0.4 dB or < 4°
for –55 dB to –70 dB
< 1 dB or < 6°
40 GHz to 50 GHz
for +15 dB to +5 dB
< 0.4 dB or < 4°
for +5 dB to –35 dB
< 0.2 dB or < 2°
for –35 dB to –50 dB
< 0.4 dB or < 4°
for –50 dB to –65 dB
< 1 dB or < 6°