ZVA Vector Network Analyzer 33
Direct generator/receiver access
These options permit direct access to the internal source output as well as to the
internal reference and measurement receiver inputs via front panel connectors.
Dynamic range with direct access utilizing these inputs is stated in the “Measurement
range” section. If all front panel jumper cables are directly connected between the
outputs and inputs, the following vector network analyzer specifications apply.
Front panel connectors
R&S
®
ZVA8 SMA,
female
R&S
®
ZVA24
2.92 mm, female
R&S
®
ZVA40
2.92 mm, female
R&S
®
ZVA50
1.85 mm, female
R&S
®
ZVA67
1.85 mm, female
Frequency range
R&S
®
ZVA8
300 kHz to 8 GHz
R&S
®
ZVA24
10 MHz to 24 GHz
R&S
®
ZVA40
10 MHz to 40 GHz
R&S
®
ZVA50
10 MHz to 50 GHz
R&S
®
ZVA67
10 MHz to 67 GHz
Dynamic range
R&S
®
ZVA8
300 kHz to 8 GHz
is reduced by 2 dB
R&S
®
ZVA24
10 MHz to 13 GHz
is reduced by 2 dB
13 GHz to 24 GHz
is reduced by 4 dB
R&S
®
ZVA40
10 MHz to 13 GHz
is reduced by 2 dB
13 GHz to 24 GHz
is reduced by 4 dB
24 GHz to 40 GHz
is reduced by 6 dB
R&S
®
ZVA50
10 MHz to 13 GHz
is reduced by 2 dB
13 GHz to 24 GHz
is reduced by 4 dB
24 GHz to 50 GHz
is reduced by 6 dB
R&S
®
ZVA67
10 MHz to 13 GHz
is reduced by 2 dB
13 GHz to 24 GHz
is reduced by 4 dB
24 GHz to 67 GHz
is reduced by 6 dB
Power range
R&S
®
ZVA8
300 kHz to 8 GHz
upper limit is reduced by 1 dB
R&S
®
ZVA24
10 MHz to 13 GHz
upper limit is reduced by 1 dB
13 GHz to 24 GHz
upper limit is reduced by 2 dB
R&S
®
ZVA40
10 MHz to 13 GHz
upper limit is reduced by 1 dB
13 GHz to 24 GHz
upper limit is reduced by 2 dB
24 GHz to 40 GHz
upper limit is reduced by 3 dB
R&S
®
ZVA50
10 MHz to 13 GHz
upper limit is reduced by 1 dB
13 GHz to 24 GHz
upper limit is reduced by 2 dB
24 GHz to 50 GHz
upper limit is reduced by 3 dB
R&S
®
ZVA67
10 MHz to 13 GHz
upper limit is reduced by 1 dB
13 GHz to 24 GHz
upper limit is reduced by 2 dB
24 GHz to 67 GHz
upper limit is reduced by 3 dB
Match R&S
®
ZVA40
10 MHz to 4 GHz
is reduced by 2 dB