General data – Atec Rohde-Schwarz-ZVA Series User Manual
Page 35
![background image](/manuals/319342/35/background.png)
Version 10.00, May 2012
Rohde & Schwarz
R&S
®
ZVA Vector Network Analyzer 35
General data
Temperature loading
in line with IEC 60068-2-1 and IEC 60068-2-2
operating temperature range
+5 °C to +40 °C
permissible temperature range
+5 °C to +40 °C
storage temperature range
–40 °C to +70 °C
Damp heat
+40 °C at 95 % rel. humidity,
in line with IEC 60068-2-30
Mechanical resistance
vibration, sinusoidal
5 Hz to 150 Hz,
in line with IEC 60068-2-6
vibration, random
10 Hz to 300 Hz,
in line with IEC 60068-2-64
shock
40 g shock spectrum,
in line with IEC 60068-2-27, MIL-STD-810
Calibration interval
1 year
EMC, RF emission
in line with EN 55011 class A, operation is
not covered in residential, commercial,
and business areas nor in small-size
companies. Thus, the instrument must not
be operated in residential, commercial,
and business areas nor in small-size
companies unless additional measures are
taken to ensure that EN 55011 class B is
met.
in line with CISPR 11/EN 55011 group 1
class A (for a shielded test setup)
The instrument complies with the emission
requirements stipulated by EN 55011 and
EN 61326-1 class A. This means that the
instrument is suitable for use in industrial
environments.
EMC, immunity
in line with IEC/EN 61326-1,
immunity for industrial environments
(excluding operating frequency)
Safety
in line with IEC 61010-1, EN 61010-1, and
UL 3111-1
Power supply
100 V to 240 V (AC) with tolerance ±10 %,
50 Hz to 60 Hz with tolerance ±5 %,
safety class I to VDE 411
Power consumption
R&S
®
ZVA8, R&S
®
ZVA24, R&S
®
ZVA40,
and R&S
®
ZVA50
450 W, typ. 310 W (standby: typ. 10 W)
R&S
®
ZVA67 only
650 W, typ. 450 W (standby: typ. 10 W)
Test mark
VDE, GS, CSA, CSA-NRTL/C,
CE conformity mark
Dimensions (W × H × D)
465.1 mm × 286.2 mm × 495.0 mm
(18.31 in × 11.27 in × 19.49 in)
Weight
25 kg (55 lb)
Shipping weight
37 kg (82 lb)