Atec Rohde-Schwarz-ZVA Series User Manual
Page 30
![background image](/manuals/319342/30/background.png)
Version 10.00, May 2012
30
Rohde & Schwarz
R&S
®
ZVA Vector Network Analyzer
DC MEAS 1 V
DC measurement input
Connector type
4-pin mini DIN, female
Voltage range
–1 V to +1 V
Measurement accuracy
2.5 % of reading + 2.5 mV
Resolution
12
bit
Bandwidth
<100
kHz
Input impedance
> 10 kΩ
Damage voltage
30 V
DC MEAS 10 V
DC measurement input
Connector type
4-pin mini DIN, female
Voltage range
–10 V to +10 V
Measurement accuracy
2.5 % of reading + 25 mV
Resolution
12
bit
Bandwidth
<100
kHz
Input impedance
> 10 kΩ
Damage voltage
30 V
PORT BIAS
DC bias input for PORT
Connector type
BNC, female
Maximum nominal input voltage
30 V
Maximum nominal input current
200 mA
Damage voltage
30 V
Damage current
500 mA
MONITOR
IBM-PC-compatible VGA monitor connector, 15-pin D-Sub (for external monitor)
USER CONTROL
several control and trigger signals, 25-pin D-Sub, 3.3 V TTL
for controlling external generators, for limit checks, sweep signals, etc.
FOOT SWITCH 1 and FOOT SWITCH 2
pin 24 and pin 25 (inputs)
control inputs
DRIVE PORT 1 to DRIVE PORT 4
pin 16 to pin 19 (outputs)
indicate driving port
CHANNEL BIT 0 to CHANNEL BIT 3
pin 8 to pin 11 (outputs) channel-specific
user-configurable bits
PASS 1 and PASS 2
pin 13 and pin 14 (outputs)
pass/fail results of limit checks
BUSY
pin 4 (output)
measurements running
READY FOR TRIGGER
pin 6 (output)
ready for trigger
EXT GEN TRIGGER
pin 21 (output)
control signal for external generator
EXT GEN BLANK
pin 22 (input)
handshake signal from external generator
EXTERNAL TRIGGER
pin 2 (input)
trigger input for analyzer
EXT TRIGGER
trigger input for analyzer
Connector type
BNC, female
TTL signal (edge-triggered)
3 V
Polarity (user-selectable)
positive or negative
Minimum pulse width
1 µs
Input impedance
> 10 kΩ