Atec Rohde-Schwarz-ZVA Series User Manual
Page 27
![background image](/manuals/319342/27/background.png)
Version 10.00, May 2012
Rohde & Schwarz
R&S
®
ZVA Vector Network Analyzer 27
Receiver linearity
referenced to –10 dBm in temperature range +18 °C to +28 °C
R&S
®
ZVA8
for +20 dB to –60 dB
50 MHz to 8 GHz
< 0.1 dB
for –60 dB to –85 dB
50 MHz to 8 GHz
typ. < 0.1 dB
R&S
®
ZVA24
for +20 dB to –30 dB
50 MHz to 700 MHz
< 0.1 dB
for –30 dB to –50 dB
50 MHz to 700 MHz
typ. < 0.1 dB
for +20 dB to +10 dB
700 MHz to 24 GHz
< 0.3 dB
for +10 dB to –45 dB
700 MHz to 24 GHz
< 0.1 dB
for –45 dB to –80 dB
700 MHz to 24 GHz
typ. < 0.1 dB
R&S
®
ZVA40
for +20 dB to –30 dB
50 MHz to 250 MHz
< 0.1 dB
for –30 dB to –50 dB
50 MHz to 250 MHz
typ. < 0.1 dB
for +10 dB to +5 dB
250 MHz to 40 GHz
< 0.3 dB
for +5 dB to –45 dB
250 MHz to 40 GHz
< 0.1 dB
for –45 dB to –65 dB
250 MHz to 40 GHz
typ. < 0.1 dB
R&S
®
ZVA50
for +20 dB to –30 dB
50 MHz to 250 MHz
< 0.1 dB
for –30 dB to –50 dB
50 MHz to 250 MHz
typ. < 0.1 dB
for +10 dB to +5 dB
250 MHz to 50 GHz
< 0.3 dB
for +5 dB to –45 dB
250 MHz to 50 GHz
< 0.1 dB
for –45 dB to –65 dB
250 MHz to 50 GHz
typ. < 0.1 dB
R&S
®
ZVA67
for +15 dB to –30 dB
50 MHz to 250 MHz
< 0.1 dB
for –30 dB to –50 dB
50 MHz to 250 MHz
typ. < 0.1 dB
for +10 dB to +5 dB
250 MHz to 67 GHz
< 0.3 dB
for +5 dB to –45 dB
250 MHz to 67 GHz
< 0.1 dB
for –45 dB to –60 dB
250 MHz to 67 GHz
typ. < 0.1 dB
Damage level
+27 dBm
Damage DC voltage
30 V