Options – Atec Rohde-Schwarz-ZVA Series User Manual
Page 31
Version 10.00, May 2012
Rohde & Schwarz
R&S
®
ZVA Vector Network Analyzer 31
Options
Generator step attenuators
R&S
®
ZVA8, R&S
®
ZVA24, and R&S
®
ZVA40:
Generator step attenuators extend the lower limit of the output power range by 70 dB.
R&S
®
ZVA50 and R&S
®
ZVA67:
Generator step attenuators extend the lower limit of the output power range by 50 dB.
Frequency range
R&S
®
ZVA8
300 kHz to 8 GHz
R&S
®
ZVA24
10 MHz to 24 GHz
R&S
®
ZVA40
10 MHz to 40 GHz
R&S
®
ZVA50
10 MHz to 50 GHz
R&S
®
ZVA67
10 MHz to 67 GHz
Power range
R&S
®
ZVA8
300 kHz to 8 GHz
upper limit is reduced by 1 dB
300 kHz to 8 GHz
lower limit is extended by 70 dB
R&S
®
ZVA24
10 MHz to 13 GHz
upper limit is reduced by 1 dB
13 GHz to 24 GHz
upper limit is reduced by 2 dB
10 MHz to 24 GHz
lower limit is extended by 70 dB
R&S
®
ZVA40
10 MHz to 13 GHz
upper limit is reduced by 1 dB
13 GHz to 24 GHz
upper limit is reduced by 2 dB
24 GHz to 40 GHz
upper limit is reduced by 3 dB
10 MHz to 40 GHz
lower limit is extended by 70 dB
R&S
®
ZVA50
10 MHz to 13 GHz
upper limit is reduced by 1 dB
13 GHz to 24 GHz
upper limit is reduced by 2 dB
24 GHz to 50 GHz
upper limit is reduced by 3 dB
10 MHz to 50 GHz
lower limit is extended by 50 dB
R&S
®
ZVA67
10 MHz to 13 GHz
upper limit is reduced by 1 dB
13 GHz to 24 GHz
upper limit is reduced by 2 dB
24 GHz to 67 GHz
upper limit is reduced by 3 dB
10 MHz to 67 GHz
lower limit is extended by 50 dB
Power accuracy
at –10 dBm without power calibration
identical to specifications
without optional step attenuators
Power linearity
(with ALC off)
R&S
®
ZVA8, R&S
®
ZVA24, and R&S
®
ZVA40
above –70 dBm
< 2 dB
from –70 dBm to –100 dBm
< 3 dB
R&S
®
ZVA50 and R&S
®
ZVA67
above –50 dBm
< 2 dB
from –50 dBm to –80 dBm
< 3 dB
Dynamic range
R&S
®
ZVA8
300 kHz to 8 GHz
is reduced by 1 dB
R&S
®
ZVA24
10 MHz to 13 GHz
is reduced by 1 dB
13 GHz to 24 GHz
is reduced by 2 dB
R&S
®
ZVA40
10 MHz to 13 GHz
is reduced by 1 dB
13 GHz to 24 GHz
is reduced by 2 dB
24 GHz to 40 GHz
is reduced by 3 dB
R&S
®
ZVA50
10 MHz to 13 GHz
is reduced by 1 dB
13 GHz to 24 GHz
is reduced by 2 dB
24 GHz to 50 GHz
is reduced by 3 dB
R&S
®
ZVA67
10 MHz to 13 GHz
is reduced by 1 dB
13 GHz to 24 GHz
is reduced by 2 dB
24 GHz to 67 GHz
is reduced by 3 dB