Atec Rohde-Schwarz-ZVA Series User Manual
Page 22
Version 10.00, May 2012
22
Rohde & Schwarz
R&S
®
ZVA Vector Network Analyzer
R&S
®
ZVA67 at –10 dBm
< 3 dB
in temperature range +18 °C to +28 °C
500 MHz to 24 GHz
24 GHz to 67 GHz
< 0.8 dB, typ. 0.3 dB
< 2 dB, typ. 1 dB
Power linearity
in temperature range +18 °C to +28 °C
(with ALC on; without power calibration)
referenced to –10 dBm
above 50 MHz
< 2 dB
R&S
®
ZVA8 above 50 MHz
< 0.8 dB, typ. 0.3 dB
R&S
®
ZVA24 above 500 MHz
< 0.8 dB, typ. 0.3 dB
R&S
®
ZVA40 above 500 MHz
< 0.8 dB, typ. 0.3 dB
R&S
®
ZVA50 above 500 MHz
< 0.8 dB, typ. 0.3 dB
R&S
®
ZVA67 above 500 MHz
< 0.8 dB, typ. 0.3 dB
Power resolution
0.01 dB
Harmonics
(output power referenced to maximum
specified output power)
R&S
®
ZVA8
300 kHz to 50 MHz at –3 dB
typ. < –30 dBc
50 MHz to 4 GHz at –5 dB
< –20 dBc, typ. < –30 dBc
4 GHz to 7 GHz at –2 dB
< –20 dBc, typ. < –30 dBc
7 GHz to 8 GHz at 0 dB
< –20 dBc, typ. < –30 dBc
R&S
®
ZVA24
10 MHz to 50 MHz at –3 dB
typ. < –30 dBc
50 MHz to 13 GHz at –3 dB
< –20 dBc, typ. < –30 dBc
13 GHz to 24 GHz at 0 dB
< –20 dBc, typ. < –30 dBc
R&S
®
ZVA40
10 MHz to 50 MHz at –3 dB
typ. < –30 dBc
50 MHz to 20 GHz at –3 dB
< –20 dBc, typ. < –30 dBc
20 GHz to 40 GHz at 0 dB
< –20 dBc, typ. < –30 dBc
R&S
®
ZVA50
10 MHz to 50 MHz at –3 dB
typ. < –30 dBc
50 MHz to 20 GHz at –3 dB
< –20 dBc, typ. < –30 dBc
20 GHz to 50 GHz at –5 dB
< –20 dBc, typ. < –30 dBc
R&S
®
ZVA67
10 MHz to 50 MHz at –3 dB
typ. < –20 dBc
50 MHz to 5 GHz at –3 dB
< –15 dBc, typ. < –25 dBc
5 GHz to 67 GHz at –3 dB
< –20 dBc, typ. < –30 dBc
Diagram: Maximum output power in dBm versus frequency in GHz of the R&S
®
ZVA8.
Diagram: Maximum output power in dBm versus frequency in GHz of the R&S
®
ZVA24.