E-series cw power sensor specifications, E-series power sensor specifications – Atec Agilent-8480 Series User Manual
Page 9
Power Linearity
Table 7
Power
Temperature (25 ºC ±5 C)
Temperature (0 to 55 ºC)
100 pW to 10 mW
±3 %
±7 %
(-70 dBm to +10 dBm)
10 mW to 100 mW
±4.5 %
±10 %
(+10 dBm to +20 dBm)
E-Series CW Power Sensor Specifications
Wide Dynamic Range CW Sensors: 100 pW to 100 mW (-70 dBm to +20 dBm).
Table 6
Model
Frequency Range
Maximum SWR
Maximum Power
Connector Type
E4412A
10 MHz – 18 GHz
*10 MHz to 30 MHz: 1.12
200 mW (+23 dBm)
Type-N (m)
30 MHz to 2 Ghz:1.15
2 GHz to 6 Ghz:1.17
6 GHz to 11 Ghz:1.2
11 GHz to 18 Ghz:1.27
E4413A
50 MHz – 26.5 GHz
50 MHz to 100 MHz: 1.21
200 mW (+23 dBm)
APC-3.5mm (m)
100 MHz to 8 Ghz:1.19
8 GHz to 18 Ghz:1.21
18 GHz to 26.5 Ghz:1.26
9
Figure 1 : Relative Mode Power Measurement Linearity with EPM series
power meter/E-series CW power sensor at 25°C±5°C (typical)
* applies to sensors with serial prefix US 3848 or greater
Power Level Used as Reference (dBm)
Power Level Being Measured (dBm)
+20
+10
+6
-20
-35
-70
-70
-35
-20
-10
+10
+20
±4%
+3%
±1%
A
B
E-series Power Sensor Specifications
The E-series of power sensors have their Calibration Factors stored in EEPROM and operate over a wide dynamic range.
They are designed for use with the EPM series of power meters and two classes of sensors are available:
• CW power sensors (E4412A and E4413A).