Power meter functions, General characteristics – Atec Agilent-8480 Series User Manual

Page 6

background image

6

Power Meter Functions

Accessed by key entry:

either

hardkeys, or softkey menu, and
programmable.

Zero:

Zeros the meter. (Power refer-

ence calibrator is switched off during
zeroing.)

Cal:

Calibrates the meter using

internal (power reference calibrator)
or external source. Reference Cal
Factor settable from 1% to 150%, in
0.1% increments.

Frequency:

Entered frequency range

is used to interpolate the calibration
factors table.

Range:

1 kHz to 999.9 GHz, settable in

1 kHz steps.

Cal Factor:

Sets the Calibration

Factor for the meter. Range: 1% to
150%, in 0.1% increments.

Relative:

Displays all successive

measurements relative to the last
displayed value.

Offset:

Allows power measurements

to be offset by –100 dB to +100 dB,
settable in 0.001dB increments, to
compensate for external loss or gain.

Save/Recall:

Store up to 10 instru-

ment states via the Save/Recall menu.

dBm/W:

Selectable units of either

Watts or dBm in absolute power;
or percent or dB for relative measure-
ments.

Filter (averaging):

Selectable from

1 to 1024. Auto-averaging provides
automatic noise compensation.

General Characteristics

Rear Panel Connectors

Recorder Outputs:

Analog 0 to

1 Volt, 1 kohm output impedance, BNC
connector. E4419B recorder outputs are
dedicated to Channel A and Channel B.

Remote Input/Output:

A TTL logic

level is output when the measurement
exceeds a predetermined limit. TTL
inputs are provided to initiate zero and
calibration cycles. RJ-45 series shielded
modular jack assembly connector. TTL
Output: high =4.8V max; low = 0.2V max
TTL

Input:

high = 3.5V min, 5V max;

low = 1V max, -0.3V min.

GPIB:

Allows communication with an

external controller.

RS-232/442:

Allows communication

with an external RS-232 or RS-422 con-
troller. Male/Plug 9 position D-subminia-
ture connector

Ground:

Binding post, accepts 4 mm

plug or bare-wire connection.

Line Power

Input Voltage Range:

85 to 264 VAC,

automatic selection.

Input Frequency Range:

50 to 440 Hz.

Power Requirement:

approximately 50

VA (14 Watts) for E4418B and E4419B.

Duty Cycle:

Duty cycle values

between 0.001% to 99.999%, in 0.001%
increments, can be entered to display
a peak power representation of meas-
ured power. The following equation is
used to calculate the displayed peak
power value: Peak Power = Measured
Power/DutyCycle.

Sensor Cal Tables:

Selects cal factor

versus frequency tables corresponding
to specified sensors ( 8480 series
only).

Limits:

High and low limits can be set

in the range –150.000 dBm to +230.000
dBm, in 0.001 dBm increments.

Preset Default Values:

dBm mode,

Rel Off, Power Reference Off, Duty
Cycle Off, Offset Off, Frequency

50 MHz, AUTO Average, Free Run,
AUTO range (for E-series sensors).

Display:

Selectable single and split

screen formats are available. A
quasi-analog display is available for
peaking measurements. The dual chan-
nel power meter can simultaneously
display any two configurations of A, B,
A/B, B/A, A-B, B-A and Relative.