E-Series E9300 Average Power Sensors

Typical E9300B/01B Power Linearity at 25 ˚C, after zero and calibration, with associated
Measurement Uncertainty

Power Range

-6 to 0 dBm

0 to +10 dBm

+10 to +20 dBm

+20 to +26 dBm

Measurement Uncertainty

± 0.65%

± 0.55%

± 0.45%

± 0.31%

Typical E9300H/01H Power Linearity at 25 ˚C, after zero and calibration, with associated
Measurement Uncertainty

Power Range

-26 to -20 dBm

-20 to –10 dBm

-10 to 0 dBm

0 to +10 dBm

+10 to +20 dBm

+20 to +26 dBm

Measurement

± 0.9%

± 0.8%

± 0.65%

± 0.55%

± 0.45%

± 0.31%

Uncertainty