Atec Agilent-8480 Series User Manual
Page 19

19
The 8480 series power sensors are
designed for use with the 435B, 436A,
437B, 438A, 70100A, E1416A and now
the E4418B and E4419B power meters.
These thermocouple and diode power
sensors provide extraordinary accuracy,
stability, and SWR over a wide range of
frequencies (100 kHz to 110 GHz) and
power levels -70 dBm to +44 dBm).
Table 14
Root Sum of Squares (rss)
uncertainty on the calibration
factor data printed on the power
sensor.
8480 Series Power Sensors (with EPM series power meters)
Freq (MHz)
8482A
8482B
8482H
8483A
0.1
1.3
2.8
1.6
1.5
0.3
1.2
2.8
1.6
1.4
1
1.2
2.8
1.6
1.4
3
1.2
2.8
1.6
1.4
10
1.3
2.8
1.6
1.6
30
1.4
2.8
1.7
1.6
50
0 (ref)
2.7
0 (ref)
0 (ref)
100
1.6
3.3
1.9
2
300
1.6
3.3
1.9
2
1000
1.4
3.3
1.7
2
2000
1.4
3.3
1.7
2.1
4000
1.5
3.1
1.8
-
Freq (GHz)
8481A
8481B
8481H
8481D
8485A
8385D
8487A
8487D
1
1.6
3
1.9
1.9
1.6
1.8
1.6
2
2
1.4
3.1
1.7
1.8
1.6
1.8
1.6
2
4
1.5
3.1
1.8
1.8
1.7
1.8
1.6
2
6
1.5
3.1
1.8
1.8
1.8
2.1
1.7
2.3
8
1.7
3.2
2
2
1.9
2.2
1.8
2.3
10
1.9
3.3
2.2
2.2
2
2.1
1.8
2.3
12
2.1
4.1
2.4
2.8
2
2.2
1.9
2.3
14
2.6
4.1
2.8
3.2
2.2
2.2
2.1
2.8
16
2.9
4.2
3
3.4
2.3
2.5
2.2
2.8
18
3.2
4.3
3.1
3.7
2
2.6
2.3
2.8
22
-
-
-
-
2.1
2
1.8
2.8
26.5
-
-
-
-
2.1
2.3
2.1
2.8
Freq (GHz)
R8486A
Q8486A
R8486D
Q8486D
8487A
8487D
26.5
2.2
-
3
-
2.1
2.8
28
2.4
-
3.2
-
2.3
3
30
2.5
-
3
-
2.1
3
33
2.1
2.8
3
4.2
2.3
3
34.5
2.1
2.8
3
4.2
2.1
3
37
2.2
2.8
3
4.2
2.3
3
40
2.2
2.9
3
4.2
2.6
3
42
-
3.9
-
4.9
3.2
2.9
44
-
3.9
-
5.1
3.6
2.9
46
-
3.9
-
5.5
4.1
3.1
48
-
4.9
-
5.8
4.5
4.5
50
-
5.3
-
6.2
5
4.5