Table 1
Accuracy
Instrumentation
Absolute:
±0.02 dB (Log) or ±0.5%
(linear). Please add the corresponding
power sensor linearity percentage
from Tables 7 and 10 (for the E-series
sensors) and Table 15 (for the 8480
series sensors).
Relative
: ±0.04 dB (Log) or ±1.0%
(linear). Please add the corresponding
power sensor linearity percentage
from Table 16 (for the 8480 series sen-
sors).
Zero Set (digital settability of
zero):
sensor dependent (refer to
Table 1). For E-series sensors, this
specification applies to a ZERO per-
formed when the sensor input is not
connected to the POWER REF.
Power Reference
Power Output:
1.00 mW (0.0 dBm).
Factory set to ±0.7% traceable to the
U.S. National Institute of Standards
and Technology.
Accuracy
: ±1.2% worst case
(±0.9% rss) for one year.
Specifications
describe the
instrument’s warranted performance
and apply after a 30 minute warm-up.
These specifications are valid over its
operating/environmental range unless
otherwise stated and after performing
a zero and calibration procedure.
Supplemental characteristics
(shown in italics) are intended to
provide additional information, useful
in applying the instrument by giving
typical (expected), but not warranted
performance parameters. These
characteristics are shown in italics or
labeled as "typical", "nominal" or
"approximate".
For information on measurement
uncertainty calculations, refer to
Application Note 64-1B,
"Fundamentals of RF and Microwave
Power Measurements", literature num-
ber 5965-6380E.
Frequency Range:
9 kHz to 110
GHz, sensor dependent.
Power Range:
–70 dBm to +44 dBm
(100 pW to 25 W), sensor dependent.
Power Sensors:
Compatible with all
8480 series and E-series sensors.
Single Sensor Dynamic Range:
90 dB maximum (E-series sensors),
50 dB maximum ( 8480 series sensors).
Display Units:
Absolute: Watts or dBm.
Relative: Percent or dB.
Display Resolution:
Selectable
resolution of 1.0, 0.1, 0.01, and 0.001
dB in Log mode, or 1 to 4 digits in
Linear mode.
Default Resolution:
0.01 dB in Log
mode, 3 digits in Linear mode.
EPM Series E4418B and E4419B Power Meters
1. Specification applies to the Low Power Path 15% to 75%
relative humidity.
Model
ZERO SET
8481A, 8482A,
±50 nW
8483A, 8485A,
8487A, R8486A, Q8486A,
8481B, 8482B
±50
µ
W
8481D, 8485D,
±20 pW
8487D
8481H, 8482H
±5
µ
W
R8486D, Q8486D
±30 pW
V8486A, W8486A
±200 nW
E4412A, E4413A
±50 pW
E9300A, E9301A,E9304A
1
±500 pW
E9300B, E9301B
1
±500 nW
E9300H, E9301H
1
±5 nW