beautypg.com

E-series e9300 average power sensors – Atec Agilent-8480 Series User Manual

Page 14

background image

14

Power Linearity (after zero and cal at ambient environmental conditions)

Table 10

Sensor

Power

Linearity

Linearity

(25 ˚C ± 10 ˚C)

(0 – 55 ˚C)

E9300A, E9301A

-60 to –10 dBm

±3.0%

±3.5%

E9304A

-10 to 0 dBm

±2.5%

±3.0%

0 to +20 dBm

±2.0%

±2.5%

E9300B, E9301B

-30 to +20 dBm

±3.5%

±4.0%

+20 to +30 dBm

±3.0%

±3.5%

+30 to +44 dBm

±2.5%

±3.0%

E9300H, E9301H

-50 to 0 dBm

±4.0%

±5.0%

0 to +10 dBm

±3.5%

±4.0%

+10 to +30 dBm

±3.0%

±3.5%

E-Series E9300 Average Power Sensors

0.5

Power (dBm)

% Error

0.4

0.3
0.2

0.1

0

-0.1

-0.2

-0.3

-0.4

-0.5

-30

-26

-22

-18

-14

-10

-6

-2

2

6

10

14

18

20

Typical E9300A/01A/04A Power Linearity at 25 ˚C, after zero and calibration, with associated
Measurement Uncertainty

Power Range

-30 to -20 dBm

-20 to -10 dBm

-10 to 0 dBm

0 to +10 dBm

+10 to +20 dBm

Measurement

±0.9%

±0.8%

±0.65%

±0.55%

±0.45%

Uncertainty