Corrected system performance with calibration kit, Specifi cation (db) – Atec Agilent-E5071C User Manual
Page 7
7
Corrected system performance with calibration kit
Option 230/235/240/245/260/265/280/285/430/435/440/445/460/465/480/485
Table 3.
Corrected system performance with type-N device connectors,
85032F calibration kit
Network analyzer:
E5071C
Calibration kit:
85032F (Type-N, 50 Ω)
Calibration:
full 2-port
IF bandwidth = 10 Hz, no averaging applied to data, environmental temperature = 23 °C
(±5 °C ) with < 1 °C deviation from calibration temperature, isolation calibration performed
Description
Specifi cation (dB)
9 kHz to
300 kHz
300 kHz to
10 MHz
10 MHz to
3 GHz
3 GHz to
6 GHz
6 GHz to
8.5 GHz
Directivity
49
49
46
40
38
Source match
41
41
40
36
35
Load match
49
49
46
40
37
Refl ection tracking
±0.011
±0.011
±0.021
±0.032
±0.054
Transmission tracking
±0.027
±0.015
±0.018
±0.056
±0.088
Transmission uncertainty (specification)
Reflection uncertainty (specification)
0.01
0.1
1
10
-
90
-
80
-
70
-
60
-
50
-
40
-
30
-
20
-
10
0
10
S11 = S22 = 0; Cal power = -10 dBm; Meas power = -10 dBm
9 kH z to 3 0 0 kH z
3 0 0 K H z to 1 0 M H z
1 0 M H z to 3 G H z
3 G H z to 6 G H z
6 G H z to 8 .5 G H z
Magnitude
Transmission coefficient (dB)
Uncertainty (dB)
Transmission coefficient (dB)
0.1
1
10
100
-90
-80
-70
-60
-50
-40
-30
-20
-10
0
10
S11 = S22 = 0; Cal power = -10 dBm; Meas power = -10 dBm
Phase
Uncertainty (Degrees)
9 kH z to 3 0 0 kH z
3 0 0 K H z to 1 0 M H z
1 0 M H z to 3 G H z
3 G H z to 6 G H z
6 G H z to 8 .5 G H z
Reflection coefficient (linear)
Uncertainty (linear)
Magnitude
0
0.01
0.02
0.03
0.04
0.05
0
0.2
0.4
0.6
0.8
1
9 KHz to 300 KHz, 300 KHz to 10 MHz
10 MHz to 3 GHz
3 GHz to 6 GHz
6 GHz to 8.5 GHz
S21 = S12 = 0; Cal power = -10 dBm; Meas power = -10 dBm
Reflection coefficient (linear)
Uncertainty (degrees)
Phase
0
2
4
6
8
10
0
0.2
0.4
0.6
0.8
1
S21 = S12 = 0; Cal power = -10 dBm; Meas power = -10 dBm
9 KHz to 300 KHz, 300 KHz to 10 MHz
10 MHz to 3 GHz
3 GHz to 6 GHz
6 GHz to 8.5 GHz