Corrected System Performance
The specifications in this section apply to measurements made with the
Agilent E5071C network analyzer under the following conditions:
•
No averaging applied to data
•
Environmental temperature of 23 °C (±5 °C) with less than 1 °C deviation
from the calibration temperature
•
Response and isolation calibration performed
System dynamic range
Table 1.
Option 230/235/240/245/260/265/280/285/430/435/440/445/
460/465/480/485
Description
Specification
SPD
System dynamic range
1, 2
9 kHz to 300 kHz
72 dB
300 kHz to 10 MHz
IF bandwidth = 3 kHz
82
dB
10 MHz to 6 GHz
98 dB
6 GHz to 8.5 GHz
92 dB
9 kHz to 300 kHz
97 dB
115 dB
300 kHz to 10 MHz
107 dB
115 dB
10 MHz to 6 GHz
IF bandwidth = 10 Hz
123 dB
130 dB
6 GHz to 7 GHz
117 dB
128 dB
7 GHz to 8 GHz
117 dB
126 dB
8 GHz to 8.5 GHz
117 dB
124 dB
Figure 1. System dynamic range (specification and actual measurement data example,
IF bandwidth 10 Hz)
80
90
100
110
120
130
140
150
160
1.E+05
1.E+09
2.E+09
3.E+09
4.E+09
5.E+09
6.E+09
7.E+09
8.E+09
Test frequency [Hz]
System dynamic range [dB]
E5071C specification
1. The test port dynamic range is calculated as the difference between the test port rms noise fl oor
and the source maximum output power. The effective dynamic range must take measurement
uncertainty and interfering signals into account.
2. The specifi cation might not be met at 5 MHz or 50 MHz.