Test port output power
7
(continued)
Table 13.
Option 2D5/2K5/4D5/4K5
Description
Specifi cation
Typical
Nominal power (preset power)
–5 dBm
Level accuracy
6
(stepped sweep mode)
1
300 kHz to 1 MHz
1 MHz to 5 MHz
5 MHz to 8.5 GHz
8.5 GHz to 20 GHz
±0.650 dB (at –5 dBm, 50 MHz absolute)
+2.0 dB, –6.0 dB
±2.0 dB
±1.0 dB
±2.5 dB
Level accuracy
(swept sweep mode)
2
300 kHz to 1 GHz
1 GHz to 8.5 GHz
8.5 GHz to 20 GHz
±5.0 dB
±2.5 dB
+5.0 dB, –7.0 dB
Level linearity
5, 6
(stepped sweep mode)
3
300 kHz to 1 MHz
1 MHz to 6 GHz
6 GHz to 8 GHz
8 GHz to 10.5 GHz
10.5 GHz to 15 GHz
15 GHz to 20 GHz
±0.75 dB (–25 to 8 dBm)
±0.75 dB (–25 to 10 dBm)
±0.75 dB (–25 to 9 dBm )
±0.75 dB (–25 to 7 dBm )
±0.75 dB (–25 to 3 dBm )
±0.75 dB (–25 to 0 dBm )
Level linearity
5
(swept sweep mode)
3
300 kHz to 1 MHz
1 MHz to 6 GHz
6 GHz to 8 GHz
8 GHz to 10.5 GHz
10.5 GHz to 15 GHz
15 GHz to 20 GHz
±1.5 dB (–25 to 8 dBm)
±1.5 dB (–25 to 10 dBm)
±1.5 dB (–25 to 9 dBm )
±1.5 dB (–25 to 7 dBm )
±1.5 dB (–25 to 3 dBm )
±1.5 dB (–25 to 0 dBm )
Range
5, 6
300 kHz to 1 MHz
1 MHz to 6 GHz
6 GHz to 8 GHz
8 GHz to 10.5 GHz
10.5 GHz to 15 GHz
15 GHz to 20 GHz
–85 to 8 dBm
–85 to 10 dBm
–85 to 9 dBm
–85 to 7 dBm
–85 to 3 dBm
–85 to 0 dBm
Sweep range
4, 5, 6
300 kHz to 1 MHz
1 MHz to 6 GHz
6 GHz to 8 GHz
8 GHz to 10.5 GHz
10.5 GHz to 15 GHz
15 GHz to 20 GHz
(Source attenuator = 0 dB)
–25 to 8 dBm
–25 to 10 dBm
–25 to 9 dBm
–25 to 7 dBm
–25 to 3 dBm
–25 to 0 dBm
Level resolution
0.05 dB
1. Level accuracy is taken at –5 dBm, relative to 50 MHz reference unless otherwise stated.
2. Level accuracy is taken at –5 dBm, relative to 50 MHz reference.
3. Level linearity given is relative to –5 dBm.
4. The sweep range shifts based on the selected source attenuator value (0 dB to 60 dB, 10 dB step).
5. The level accuracy specifi cation needs to be taken into account for test port output power level.
6. Power calibration using an external power meter improves level accuracy of the test port output power.
Proper power meters/sensors, and the 82357B USB-GPIB interface are required to conduct power calibration.
7. Source output performance on port 1 only. Other port output performance is typical.