Specifi cation (db) – Atec Agilent-E5071C User Manual
Page 11
11
Option 2D5/2K5/4D5/4K5
Table 7.
Corrected system performance with 3.5 mm device connectors,
85052D calibration kit
Network analyzer:
E5071C
Calibration kit:
85052D (3.5 mm, 50 Ω)
Calibration:
full 2-port
IF bandwidth = 10 Hz, no averaging applied to data, environmental temperature = 23 °C
(±5 °C ) with < 1 °C deviation from calibration temperature, isolation calibration performed
Description
Specifi cation (dB)
300 kHz to
500 MHz
500 MHz to
2 GHz
2 GHz to
6 GHz
6 GHz to
20 GHz
Directivity
42
42
38
36
Source match
37
37
31
28
Load match
42
42
38
36
Refl ection tracking
±0.003
±0.003
±0.004
±0.008
Transmission tracking
±0.068
±0.034
±0.100
±0.208
Transmission uncertainty (specification)
Reflection uncertainty (specification)
0.01
0.1
1
10
-
90
-
80
-
70
-
60
-
50
-
40
-
30
-
20
-
10
0
10
3 0 0 K Hz to 5 0 0 M Hz
5 0 0 M Hz to 2 G Hz
2 G Hz to 6 G Hz
6 G Hz to 2 0 G Hz
S11 = S22 = 0; Cal power = -10 dBm; Meas power = -10 dBm
U
n
c
e
rta
in
ty
(d
B
)
Transmission Coefficient (dB)
Magnitude
0.1
1
10
100
-90
-80
-70
-60
-50
-40
-30
-20
-10
0
10
3 0 0 K Hz to 5 0 0 M Hz
5 0 0 M Hz to 2 G Hz
2 G Hz to 6 G Hz
6 G Hz to 2 0 G Hz
S11 = S22 = 0; Cal power = -10 dBm; Meas power = -10 dBm
Phase
U
ncer
tai
nt
y (
D
eg
rees)
Trans mis s ion C oefficient (dB )
0
0 .0 1
0 .0 2
0 .0 3
0 .0 4
0 .0 5
0
0 .2
0 .4
0 .6
0 .8
1
U
n
cer
tai
n
ty (
lin
ear
)
Reflection coefficient (linear)
Magnitude
3 0 0 K H z to 5 0 0 M H z, 5 0 0 M H z to 2 G H z
2 G H z to 6 G H z
6 G H z to 2 0 G H z
S 2 1 = S 1 2 = 0 ; Cal power = -10 dBm; Meas power = -10 dBm
0
2
4
6
8
1 0
0
0 .2
0 .4
0 .6
0 .8
1
U
n
ce
rta
in
ty
(d
eg
re
es
)
Reflection coefficient (linear)
Phase
3 0 0 K H z to 5 0 0 M H z, 5 0 0 M H z to 2 G H z
2 G H z to 6 G H z
6 G H z to 2 0 G H z
S 2 1 = S 1 2 = 0 ; Cal power = -10 dBm; Meas power = -10 dBm