Test port input (group delay)

1

Table 26.

Option 230/235/240/245/260/265/280/285/430/435/440/

445/460/465/480/485

Description

Specifi cation

Supplemental information

Aperture (selectable)

(frequency span)/(number
of points - 1)

Maximum aperture

25% of frequency span

Minimum delay

Limited to measuring no more
than 180° of phase change
within the minimum aperture.

Accuracy

See graph below, typical

The following graph shows group delay accuracy with type-N connectors, full 2-port
calibration and a 10 Hz IF bandwidth.

Calibration kit (85032F). Insertion loss is assumed to be < 2 dB.

In general, the following formula can be used to determine the accuracy, in seconds, of a
specifi c group delay measurement:

±phase accuracy (degrees) / [360 x aperture (Hz)]

1. Group delay is computed by measuring the phase change within a specifi ed step (determined by

the frequency span and the number of points per sweep).

Accuracy (nSec)

Aperture (MHz)

0.01 0.1 1 10 100

100

10

1

0.1

0.01

0.001